Objectives: To compare audiometric results and complications of stapes surgery with two different types of piston prosthesis, the Portmann Clip Piston (Medtronic) (PCP) and the Soft Clip Piston (Kurz) (SCP).

Material And Methods: Study conducted on 64 patients who underwent primary stapedotomy from 2008 to 2011. We matched for each case of stapedotomy with the PCP (Medtronic Xomed Inc. Portmann Clip Piston Stainless Steel/Fluoroplastic) a case with the SCP (Heinz Kurz GmbH Medizintechnik Soft Piston Clip Titanium). Each group consisted of 32 patients, and patients in both groups were matched with respect to gender, age, bilateral or unilateral otosclerosis, otological symptoms (tinnitus, vertigo or dizziness), family history, operated side and the Portmann grading for otosclerosis. The length of the prosthesis used was reported. Post-operative complications such as tinnitus, vertigo, hearing loss and altered taste were documented. Each patient was subjected to a preoperative and postoperative audiogram (follow-up at the second month after the surgery). We used the Student test for statistical analysis. Statistical significance was set at < 0.01.

Results: None of the patients experienced a post-operative hearing loss and none required a later revision surgery. No statistically significant difference was found between the two populations regarding demographic data (age, sex, side, bilaterality, family history, stage and lenght of piston) and hearing level (> 0.01) in the air, bone conduction and air-bone gap (ABG). Postoperative complications did not result to be significantly different between the two groups. Also, both groups showed a significant improvement (< 0.01) in the post-operative air, bone conduction and air-bone gap. There was no statistically significant difference (> 0.01) between the post-operative hearing results (bone conduction, air conduction, air-bone gap) using the two pistons. The mean ABG improvement was respectively 16.63 dB in the SCP group and 20.59 dB in the PCP group.

Conclusion: The titanium Soft clip piston (SCP) is a good alternative to the Portmann clip piston (PCP). Nevertheless there are some differences in the surgical fixing of these two pistons in the correct position.

Download full-text PDF

Source

Publication Analysis

Top Keywords

clip piston
20
portmann clip
12
bone conduction
12
conduction air-bone
12
air-bone gap
12
stapes surgery
8
piston
8
soft clip
8
tinnitus vertigo
8
family history
8

Similar Publications

 Stapes surgery for otosclerosis is a precise surgical procedure. To achieve excellent hearing results, a firm and stable attachment of the prosthesis to the long process of incus (LPI) is necessary. The present study provides details on the attachment site in two dimensions to choose an appropriate prosthesis and to ensure firm attachment for better surgical outcomes.

View Article and Find Full Text PDF

Audiological outcome after stapes surgery in relation to prosthesis type.

Eur Arch Otorhinolaryngol

July 2023

Department of Otorhinolaryngology, Head, Neck, Plastic and Cosmetic Surgery, Friedrichshain Clinic, Vivantes Hearing Center (HZB), Landsberger Allee 49, 10249, Berlin, Germany.

Purpose: Different techniques are used to fix crimp and CliP® Piston stapes prostheses to the long process of the incus (LPI). The CliP® Piston provides a stiff connection in contrast to the static bended loop of the crimp prosthesis, which imitates the physiological incudostapedial joint (ISJ) and thereby potentially leads to different hearing outcome.

Methods: In a retrospective single-center study of German-speaking one hundred and ninety patients who underwent stapes surgery CliP® Piston or crimp prostheses between the years of 2016 and 2019 by the same surgeon and in the same setting.

View Article and Find Full Text PDF

A New Stapes-Head Coupler for the Vibrant Soundbridge System.

Audiol Neurootol

September 2021

Department of Otolaryngology, Head and Neck Surgery, University Hospital Zürich, Zürich, Switzerland,

Introduction: The Vibrant Soundbridge (MED-EL Medical Electronics, Austria) is an active middle ear implant with a floating mass transducer (FMT) for patients with conductive, sensorineural, or mixed hearing loss. While the FMT is vertically aligned above the stapes head (SH) with the current Vibroplasty Clip coupler (MED-EL Medical Electronics), the new SH coupler was developed to mount the FMT on the inferior side of the stapes and to fit in the reduced middle ear space after canal-wall-down mastoidectomy.

Methods: Using 11 human cadaveric temporal bones (TBs), placements of the new SH couplers on the stapes were examined, and effective stimuli to the cochlea were evaluated by measuring piston-like motion of the stapes footplate with a current of 1 mA on the FMT.

View Article and Find Full Text PDF

The aim of the current study was to investigate the construction of the bone bridge and tibial plateau under arthroscopy during meniscal allograft transplantation, in order to simplify and enhance the accuracy of bone bridge fixation intraoperatively. A traction line passed through the attachment of the anterior and posterior horns of the superior meniscus to the bone bridge was used to pull the bone bridge into the knee joint cavity and fix the anteroposterior horns of the meniscus. At the junction of the body of the meniscus and the posterior and anterior horns of the meniscus, a traction line was created at the anterior and posterior 1/3 of the meniscus to pull and fix the meniscus.

View Article and Find Full Text PDF

High Speed Pneumatic Stepper Motor for MRI Applications.

Ann Biomed Eng

March 2019

School of Electrical and Computer Engineering, University of Georgia, Athens, GA, USA.

Presented in this paper is an exploration into the efficacy of a plastic, four-cylinder piston pneumatic motor designed for driving medical instruments inside magnetic resonance imaging (MRI) systems. Because of the increasing use of MRI for diagnostic and interventional purposes and the benefits that could be realized by operating with real-time MR image guidance, there exists a significant need for MRI compliant surgical devices. Some type of actuation mechanism is necessary to drive such devices.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!