Analysis of iodine in food samples by inductively coupled plasma-mass spectrometry.

Food Addit Contam Part A Chem Anal Control Expo Risk Assess

a Center for Food Safety and Applied Nutrition , US Food and Drug Administration (USFDA) , College Park, MD , USA.

Published: October 2016

This work shows a method for the determination of iodine in a variety of food samples and reference materials using inductively coupled plasma-mass spectrometry (ICP-MS) following alkaline extraction. Optimisation of the addition of organic carbon showed that a minimum of 3% 2-propanol was necessary for a constant ratio of iodine to internal standard. The limit of quantification (LOQ), calculated as 30σ for the method, was 36 ng g(-1) in solid food samples. For method validation, seven standard reference materials (SRM) and 21 fortified food samples were used. The precision (%RSD) of the measurements was in the 2-7% range. Accuracies for the SRMs were 85-105%, while the fortified food samples showed 81-119% recoveries, including a number of samples fortified at 50% of the LOQ.