Objective: To evaluate the accuracy and safety of pedicle screw insertion with the aid of novel patient-specific drill-guide templates in scoliosis cases.

Methods: Ten patients with scoliosis were selected to participate in the research (the observation group) from December 2013 to December 2014. The data was obtained from CT scanning, and put into the computer to perform reconstruction of spine, simulation of pedicle screw insertion, and design of patient-specific drill-guide templates with software. The templates were made with rapid prototyping technique. After sterilization, the templates were used to aid the pedicle screw insertion intraoperatively. The blood loss, operation duration, change of creatinine level pre- and post-operation, and complications related to pedicle screw insertion were recorded. The location of pedicle screws were graded so as to evaluate the accuracy. A comparative study was then performed with the data of ten scoliosis cases operated with free-hand method during the same period (control group). There were 5 cases of idiopathic scoliosis and 5 cases of congenital scoliosis in the observation group, including 3 males and 7 females. Their average age was 11.9 years old (ranged, 4 to 18 years old), and the average Cobb angle of main curve was 54.9° (ranged, 42.1° to 78.4°). There were also 5 cases of idiopathic scoliosis and 5 cases of congenital scoliosis in the control group,including 2 males and 8 females. Their average age was 12.6 years old (ranged, 6 to 17 years old), and the average Cobb angle of main curve was 56.6° (ranged, 38.2° to 93.4°).

Results: A total of 167 pedicle screws were inserted intraoperatively, with 138 screws (82.6%) in grade I, 26 screws (15.0%) in grade II, 4 screws in grade III (2.4%), but no screws in grade IV according to the CT image. There were 29 (17.4%) screws perforated, and 163 (97.6%) screws could be accepted. In the control group, a total of 165 pedicle screws were inserted intraoperatively, with 98 screws (59.4%) in grade I, 39 screws (23.6%) in grade II, 21 screws in grade III (12.7%), and 7 screws in grade IV (4.2%). There were 67 (40.6%) screws perforated, and 137 (83.0%) screws could be accepted. The grade distribution of screw position, ratio of perforated and accepted screws were significantly different between the two groups respectively (Z=-5.013, P=0.000; χ2=9.347, P=0.002; χ2=20.242, P=0.000). The correction rate of Cobb angle were (74.1±10.0)% vs (69.7±17.6)%; blood loss were (455±447) ml vs (415±389) ml; operation duration were (163.5±53.7) min vs (164.0±48.7) min; and the changes of creatinine level pre- and post-operatively were (-5.3±3.2) μmol/L vs (-3.4±3.1) μmol/L; all above data had no significant differences respectively (t=0.696, P=0.496; t=0.214, P=0.833; t=0.022, P=0.983; t=1.375, P=0.192). There were no complications related to pedicle screw insertion in each group.

Conclusion: The novel patient-specific drill guide template can be used to assist the insertion of pedicle screws in scoliosis cases with much higher accuracy than that of freehand method and fair safety.

Download full-text PDF

Source

Publication Analysis

Top Keywords

pedicle screw
20
screw insertion
20
screws
17
pedicle screws
16
scoliosis cases
16
grade screws
16
screws grade
16
patient-specific drill-guide
12
drill-guide templates
12
cobb angle
12

Similar Publications

Background: Pedicle screw insertion in posterior spinal surgery can cause vascular injuries, including rare intercostal artery pseudoaneurysms, which are typically discovered incidentally during reimaging. Onyx embolization is an effective treatment for small artery pseudoaneurysms.

Observations: A 36-year-old man who had initially presented with back pain that remained unresponsive to nonsteroidal anti-inflammatory drugs was diagnosed with a T7-8 sarcomatous lesion confirmed by magnetic resonance imaging and biopsy.

View Article and Find Full Text PDF

Background: Recent advances within the last decade have allowed robotics to become commonplace in the operating room. In the field of neurosurgery, robotics assist surgeons in pedicle screw placement and vertebral fusion procedures. The purpose of this review is to look at currently used spinal robots available on the market and compare their overall accuracy, cost, radiation exposure, general adverse events, and hospital readmission rates.

View Article and Find Full Text PDF

Preserving Cervical Mobility: A Novel Robot-Assisted Approach for Atlas Fracture Fixation.

Am J Case Rep

January 2025

Department of Orthopedic Surgery, Sichuan Provincial People's Hospital, School of Medicine, University of Electronic Science and Technology of China, Chengdu, Sichuan, China.

BACKGROUND The management of unstable atlas fractures remains a subject of ongoing debate and controversy. The conservative surgical treatment commonly involves fusion, resulting in severe loss of cervical spine mobility, and a large incisions and extensive tissue dissection are required. We aim to introduce a novel concept and surgical approach for treating atlas fracture, one that involves minimizing trauma while maintaining mobility of the upper cervical spine without resorting to fusion.

View Article and Find Full Text PDF

Objective: To observe the clinical outcomes of anterior approach for the revision surgery following unsuccessful bone cement augmentation in osteoporotic vertebral compression fractures.

Methods: A total of 10 patients who experienced unsuccessful bone cement augmentation underwent anterior revision surgery between January 2020 and December 2021. There were 2 males and 8 females.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!