Purpose: To establish the extent of dose reduction that can be achieved using a computed tomography (CT) quality assurance phantom with various user-defined protocols at different milliamperage (mA) settings and kilovoltage peak (kVp) without affecting image quality.

Methods: Images of a CT quality assurance phantom containing line-pair patterns and 3 tissue-equivalent inserts in water were acquired with different mA settings on 4 different CT scanners. The kVp was kept constant at 120 while the mA was changed in increments of 20, from 30 mA to 190 mA. Three images from each series acquired at different mA settings were selected: one each for the quantitative analysis of spatial resolution, contrast-to-noise ratio (CNR), and image noise using image analysis freeware.

Results: Statistical analysis was carried out with a statistical probability of P < .05 indicating significant difference. Results of the statistical analysis revealed significant differences in the CNR of all 3 types of tissue-equivalent inserts (P < .001) and in image noise (P < .001) as the mA was changed. However, changes in mA did not produce significant differences in the spatial resolution of the images (P > .06).

Conclusion: A radiation dose reduction of 25% was achieved in 4 different CT scanners without significantly affecting the 3 image quality parameters.

Download full-text PDF

Source

Publication Analysis

Top Keywords

image quality
8
quality parameters
8
computed tomography
8
dose reduction
8
quality assurance
8
assurance phantom
8
tissue-equivalent inserts
8
acquired settings
8
spatial resolution
8
image noise
8

Similar Publications

The quality of CT images obtained from hepatocellular carcinoma (HCC) patients is complex, affecting diagnostic accuracy, precision, and radiation dose assessment due to increased exposure risks. The study evaluated image quality qualitatively and quantitatively by comparing quality levels with an effective radiation dose to ensure acceptable quality accuracy. This study retrospectively reviewed 100 known HCC patients (Li-RADS-5) who underwent multidetector computed tomography (MDCT) multiphasic scans for follow-up of their health condition between January and October 2023.

View Article and Find Full Text PDF

The kinetics, oil migration pattern and the role of frying media during immersion frying of '', a dairy dessert, at the microstructural level were studied using confocal laser scanning microscopy (CLSM). After 6 min of frying, the depth of oil migration in increased from 0 to 3.16 mm in clarified butter (locally called '') and 3.

View Article and Find Full Text PDF

MITIGATING OVER-SATURATED FLUORESCENCE IMAGES THROUGH A SEMI-SUPERVISED GENERATIVE ADVERSARIAL NETWORK.

Proc IEEE Int Symp Biomed Imaging

May 2024

Department of Electrical and Computer Engineering, Nashville, TN, USA.

Multiplex immunofluorescence (MxIF) imaging is a critical tool in biomedical research, offering detailed insights into cell composition and spatial context. As an example, DAPI staining identifies cell nuclei, while CD20 staining helps segment cell membranes in MxIF. However, a persistent challenge in MxIF is saturation artifacts, which hinder single-cell level analysis in areas with over-saturated pixels.

View Article and Find Full Text PDF

Patient engagement and shared decision-making (SDM) between patients and clinicians is the foundation of patient-centered care. It aims to reach a treatment option that fits the patient's preference and is guideline-concordant. We sought to evaluate the possible causes and outcomes of patient's non-guideline-concordant care choices.

View Article and Find Full Text PDF

Fracture surgeries are frequently accompanied by severe pain, necessitating efficacious pain management strategies to enhance postoperative recovery. Nerve block techniques, which are critical in mitigating pain, involve the targeted administration of local anesthetics to disrupt nerve signal transmission, thereby achieving significant analgesia. Traditionally, these techniques rely on anatomical landmarks and the clinician's expertise, which can introduce variability and potential risks.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!