Background: It is known that the arterial carbon dioxide pressure (PaCO2) is useful for emergency physicians to assess the severity of dyspnoeic spontaneously breathing patients. Transcutaneous carbon dioxide pressure (PtcCO2) measurements could be a non-invasive alternative to PaCO2 measurements obtained by blood gas samples, as suggested in previous studies. This study evaluates the reliability of a new device in the emergency department (ED).
Methods: We prospectively included patients presenting to the ED with respiratory distress who were breathing spontaneously or under non-invasive ventilation. We simultaneously performed arterial blood gas measurements and measurement of PtcCO2 using a sensor placed either on the forearm or the side of the chest and connected to the TCM4 CombiM device. The agreement between PaCO2 and PtcCO2 was assessed using the Bland-Altman method.
Results: Sixty-seven spontaneously breathing patients were prospectively included (mean age 70 years, 52% men) and 64 first measurements of PtcCO2 (out of 67) were analysed out of the 97 performed. Nineteen patients (28%) had pneumonia, 19 (28%) had acute heart failure and 19 (28%) had an exacerbation of chronic obstructive pulmonary disease. Mean PaCO2 was 49 mm Hg (range 22-103). The mean difference between PaCO2 and PtcCO2 was 9 mm Hg (range -47 to +54) with 95% limits of agreement of -21.8 mm Hg and 39.7 mm Hg. Only 36.3% of the measurement differences were within 5 mm Hg.
Conclusions: Our results show that PtcCO2 measured by the TCM4 device could not replace PaCO2 obtained by arterial blood gas analysis.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1136/emermed-2015-205203 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!