Magnetic resonance-guided motorized transcranial ultrasound system for blood-brain barrier permeabilization along arbitrary trajectories in rodents.

J Ther Ultrasound

UNIRS, Neurospin, I2BM, Direction des Sciences du Vivant, Commissariat à l'Energie Atomique et aux Energies Alternatives, Bâtiment 145, 91191 Gif sur Yvette, France.

Published: December 2015

Background: Focused ultrasound combined with microbubble injection is capable of locally and transiently enhancing the permeability of the blood-brain barrier (BBB). Magnetic resonance imaging (MRI) guidance enables to plan, monitor, and characterize the BBB disruption. Being able to precisely and remotely control the permeabilization location is of great interest to perform reproducible drug delivery protocols.

Methods: In this study, we developed an MR-guided motorized focused ultrasound (FUS) system allowing the transducer displacement within preclinical MRI scanners, coupled with real-time transfer and reconstruction of MRI images, to help ultrasound guidance. Capabilities of this new device to deliver large molecules to the brain on either single locations or along arbitrary trajectories were characterized in vivo on healthy rats and mice using 1.5 MHz ultrasound sonications combined with microbubble injection. The efficacy of BBB permeabilization was assessed by injecting a gadolinium-based MR contrast agent that does not cross the intact BBB.

Results: The compact motorized FUS system developed in this work fits into the 9-cm inner diameter of the gradient insert installed on our 7-T preclinical MRI scanners. MR images acquired after contrast agent injection confirmed that this device can be used to enhance BBB permeability along remotely controlled spatial trajectories of the FUS beam in both rats and mice. The two-axis motor stage enables reaching any region of interest in the rodent brain. The positioning error when targeting the same anatomical location on different animals was estimated to be smaller than 0.5 mm. Finally, this device was demonstrated to be useful for testing BBB opening at various acoustic pressures (0.2, 0.4, 0.7, and 0.9 MPa) in the same animal and during one single ultrasound session.

Conclusions: Our system offers the unique possibility to move the transducer within a high magnetic field preclinical MRI scanner, thus enabling the delivery of large molecules to virtually any rodent brain area in a non-invasive manner. It results in time-saving and reproducibility and could be used to either deliver drugs over large parts of the brain or test different acoustic conditions on the same animal during the same session, therefore reducing physiological variability.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4690289PMC
http://dx.doi.org/10.1186/s40349-015-0044-5DOI Listing

Publication Analysis

Top Keywords

preclinical mri
12
blood-brain barrier
8
arbitrary trajectories
8
focused ultrasound
8
combined microbubble
8
microbubble injection
8
fus system
8
mri scanners
8
large molecules
8
rats mice
8

Similar Publications

Introduction: Alzheimer's disease (AD) is now diagnosed biologically. Since subjective cognitive decline (SCD) may indicate preclinical AD, assessing AD-biomarkers is crucial. We investigated cognitive and neurodegenerative trajectories in SCD over 24 months based on biomarker positivity, and evaluated the predictive value of plasma biomarkers.

View Article and Find Full Text PDF

Background And Objectives: Magnetic resonance imaging (MRI) and neurohistopathology are important correlates for evaluation of disease progression in multiple sclerosis (MS). Here we used experimental autoimmune encephalomyelitis (EAE) as an animal model of MS to determine the correlation between clinical EAE severity, MRI and histopathological parameters.

Methods: N = 11 female C57BL/6J mice were immunized with human myelin oligodendrocyte glycoprotein 1-125, while N = 9 remained non-immunized.

View Article and Find Full Text PDF

Radiation therapy (RT) is widely used to treat thoracic cancers but carries a risk of radiation-induced heart disease (RIHD). This study aimed to detect early markers of RIHD using machine learning (ML) techniques and cardiac MRI in a rat model. SS.

View Article and Find Full Text PDF

Targeting Reactive Oxygen Species for Diagnosis of Various Diseases.

J Funct Biomater

December 2024

Cardiovascular Institute, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, MA 02115, USA.

Reactive oxygen species (ROS) are generated predominantly during cellular respiration and play a significant role in signaling within the cell and between cells. However, excessive accumulation of ROS can lead to cellular dysfunction, disease progression, and apoptosis that can lead to organ dysfunction. To overcome the short half-life of ROS and the relatively small amount produced, various imaging methods have been developed, using both endogenous and exogenous means to monitor ROS in disease settings.

View Article and Find Full Text PDF

Advances in Imaging of Traumatic Nerve Injuries.

J Am Acad Orthop Surg

December 2024

From the Department of Orthopaedic Surgery, University of Utah, Salt Lake City, UT (Graesser), the Washington University School of Medicine in St. Louis, Mallinckrodt Institute of Radiology, St. Louis, MO (Parsons), and the Department of Orthopaedic Surgery, Washington University School of Medicine in St. Louis, St. Louis, MO (Olafsen, Dy, and Brogan).

Traumatic peripheral nerve injuries represent a spectrum of conditions and remain challenging to diagnose and prognosticate. High-resolution ultrasonography and magnetic resonance neurography have emerged as useful diagnostic modalities in the evaluation of traumatic peripheral nerve and brachial plexus injuries. Ultrasonography is noninvasive, is able to rapidly interrogate large areas and multiple nerves, allows for a dynamic assessment of nerves and their surrounding anatomy, and is cost-effective.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!