A 24 mm COMS-like eye plaque was developed to meet the treatment needs of our eye plaque brachytherapy practice. As part of commissioning, it was necessary to determine the new plaque's seed coordinates. The FARO Edge, a commercially available measurement arm, was chosen for this purpose. In order to validate the FARO Edge method, it was first used to measure the seed marker coordinates in the silastic molds for the standard 10, 18, and 20 mm COMS plaques, and the results were compared with the standard published Task Group 129 coordinates by a nonlinear least squares match in MATLAB version R2013a. All measured coordinates were within 0.60 mm, and root mean square deviation was 0.12, 0.23, and 0.35 mm for the 10, 18, and 20 mm molds, respectively. The FARO Edge was then used to measure the seed marker locations in the new 24 mm silastic mold. Those values were compared to the manufacturing specification coordinates and were found to demonstrate good agreement, with a maximum deviation of 0.56mm and a root mean square deviation of 0.37 mm. The FARO Edge is deemed to be a reliable method for determining seed coordinates for COMS silastics, and the seed coordinates for the new 24 mm plaque are presented.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5690999PMC
http://dx.doi.org/10.1120/jacmp.v16i6.5721DOI Listing

Publication Analysis

Top Keywords

faro edge
20
seed coordinates
16
eye plaque
8
measure seed
8
seed marker
8
root square
8
square deviation
8
coordinates
7
seed
6
faro
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!