Constancy checks of a well-type ionization chamber should be performed regularly as part of a quality assurance regime. The goal of this work was to test the feasibility of using a linear accelerator and an orthovoltage unit to check the constancy of a well-type chamber's response to an external radiation source. The reproducibility, linearity with dose, variation with dose-rate, and variation between energy-matched units of the well-type chamber response when exposed to 6 MV beams was examined. The robustness to errors in establishing the measurement conditions, including setting the source-to-surface distance and gantry angle, rotation of the chamber around the central axis of the beam, and the effect of changing the length of the chamber cable exposed to the field, were tested. The reproducibility and linearity with dose of the chamber response, and robustness to errors in establishing the measurement conditions for 100 kVp and 250 kVp beams from an orthovoltage unit, were also examined. The combined uncertainty, including contributions from errors in establishing the reference conditions, for well-type chamber measurements using a 6 MV beam from a linear accelerator is 1.0%. The combined uncertainties for measurements using 100 and 250 kVp beams were 1.8% and 1.5%, respectively. When focus-source distance errors were reduced to ≤ 1 mm, the combined uncertainties for the 100 and 250 kVp beams were 1.2% and 1.1%, respectively, when the dose to the chamber was confined to the linear region of the dose-response curve. The response of a well-type chamber should remain constant to within 1.2% when exposed to a constant dose from an external beam unit, if reference conditions can be reproducibly established. However, the uncertainty for establishing reference conditions for output measurements for an orthovoltage unit can be reduced, which would justify a reduction of the tolerance for constancy measurements.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5691027PMC
http://dx.doi.org/10.1120/jacmp.v16i6.5608DOI Listing

Publication Analysis

Top Keywords

orthovoltage unit
12
well-type chamber
12
errors establishing
12
250 kvp
12
kvp beams
12
reference conditions
12
constancy checks
8
checks well-type
8
well-type ionization
8
chamber
8

Similar Publications

Purpose: Minibeam radiation therapy (MBRT) is characterized by the delivery of submillimeter-wide regions of high "peak" and low "valley" doses throughout a tumor. Preclinical studies have long shown the promise of this technique, and we report here the first clinical implementation of MBRT.

Methods And Materials: A clinical orthovoltage unit was commissioned for MBRT patient treatments using 3-, 4-, 5-, 8-, and 10-cm diameter cones.

View Article and Find Full Text PDF

Spatially fractionated radiotherapy is showing promise as a treatment modality. Initial focus was on beams of photons at low energy produced from a synchrotron but more recently research has expanded to include applications in proton therapy. Interest in photon beams remains and this is the focus of this paperThis study presents a 3D printed tungsten minibeam collimator intended to produce peak-to-valley dose ratios (PVDR) of between seven and ten with a 1 MV, bremsstrahlung generated, photon beam.

View Article and Find Full Text PDF

. The purpose of this study is to present data from the clinical commissioning of an Xstrahl 150 x-ray unit used for superficial radiotherapy,. Commissioning tasks included vendor acceptance tests, timer reproducibility, linearity and end-effect measurements, half-value layer (HVL) measurements, inverse square law verification, head-leakage measurements, and beam output calibration.

View Article and Find Full Text PDF

Kilovoltage radiotherapy dose calculations are generally performed with manual point dose calculations based on water dosimetry. Tissue heterogeneities, irregular surfaces, and introduction of lead cutouts for treatment are either not taken into account or crudely approximated in manual calculations. Full Monte Carlo (MC) simulations can account for these limitations but require a validated treatment unit model, accurately segmented patient tissues and a treatment planning interface (TPI) to facilitate the simulation setup and result analysis.

View Article and Find Full Text PDF

Relative dosimetry measurements are required to fully commission kilovoltage X-ray units for superficial and orthovoltage X-ray therapy. Validation of these relative dosimetry measurements with Monte Carlo methods is advantageous being independent of the measurement process. In this study use is made of the X-ray spectrum generating program SpekPy along with the EGSnrc Monte Carlo code to calculate depth doses and explore the dosimetry effect of changes in backscatter.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!