Purpose: To determine the specificity of the Pediatric Vision Scanner, a binocular retinal birefringence scanner, in its intended setting, a pediatric primary care office.
Methods: A total of 102 preschool children (age 2-6 years) were screened during a well-child pediatric visit using the Pediatric Vision Scanner and the SureSight Autorefractor and completed a masked comprehensive pediatric ophthalmic examination (gold standard examination).
Results: Based on the gold standard examination, one child had anisometropic amblyopia, and the remaining 101 had no amblyopia or strabismus. Specificity of the Pediatric Vision Scanner was 90% (95% CI, 82%-95%) while specificity of the SureSight was 87% (95%CI, 79-93%). Combining these results with the sensitivity of the devices determined in our previous study conducted in a pediatric ophthalmology office setting, the positive likelihood ratio for the Pediatric Vision Scanner was 10.2; for the SureSight, 5.0. The negative likelihood ratio for the Pediatric Vision Scanner was 0.03; for the SureSight, 0.42, a significant difference.
Conclusions: The Pediatric Vision Scanner had high specificity (90%) in screening for amblyopia and strabismus as part of a pediatric well-child visit. Likelihood ratio analysis suggests that affected children have a high probability of being correctly identified by the Pediatric Vision Scanner. The high level of confidence conferred by Pediatric Vision Scanner screening may remove an important barrier to vision screening in pediatric primary care.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4688569 | PMC |
http://dx.doi.org/10.1016/j.jaapos.2015.09.004 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!