Inter-laboratory evaluation of the performance parameters of a Lateral Flow Test device for the detection of Bluetongue virus-specific antibodies.

J Virol Methods

Veterinary and Agrochemical Research Centre (CODA-CERVA), Coordination of the Veterinary Diagnostic, Epidemiology and Risk Analysis Unit (CDV-ERA), Groeselenberg 99, 1180 Brussels, Belgium; Ghent University, Department of Virology, Parasitology and Immunology, Faculty of Veterinary Medicine, Salisburylaan 133, 9820 Merelbeke, Belgium.

Published: February 2016

Bluetongue (BT) is a viral vector-borne disease affecting domestic and wild ruminants worldwide. In this study, a commercial rapid immuno-chromatographic method or Lateral Flow Test (LFT) device, for the detection of BT virus-specific antibodies in animal serum, was evaluated in an international inter-laboratory proficiency test. The evaluation was done with sera samples of variable background (ruminant species, serotype, field samples, experimental infections, vaccinated animals). The diagnostic sensitivity was 100% (95% C.I. [90.5-100]) and the diagnostic specificity was 95.2% (95% C.I. [76.2-99.9]). The repeatability (accordance) and reproducibility (concordance) were 100% for seropositive samples but were lower for two of the seronegative samples (45% and 89% respectively). The analytical sensitivity, evaluated by testing positive sera at increasing dilutions was better for the BT LFT compared to some commercial ELISAs. Seroconversion of an infected sheep was detected at 4 days post infection. Analytical specificity was impaired by cross-reactions observed with some of the samples seropositive for Epizootic Haemorrhagic Disease Virus (EHDV). The agreement (Cohen's kappa) between the LFT and a commercial BT competitive ELISA was 0.79 (95% CI [0.62-0.95]). Based on these results, it can be concluded that the BT LFT device is a rapid and sensitive first-line serological test that can be used in the field, especially in areas endemic for the disease where there is a lack of diagnostic facilities.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jviromet.2015.12.001DOI Listing

Publication Analysis

Top Keywords

lateral flow
8
flow test
8
device detection
8
virus-specific antibodies
8
lft device
8
samples
5
inter-laboratory evaluation
4
evaluation performance
4
performance parameters
4
parameters lateral
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!