In magnetic resonance imaging (MRI) of brain, fast spin echo-fluid attenuated inversion recovery (FSE-FLAIR) is widely used to detect high intensity areas (HIAs). The way of getting FLAIR contrast at high speed includes echo planar imaging-FLAIR (EPI-FLAIR), and in EPI-FLAIR there are single shot EPI-FLAIR (SS-EPI-FLAIR) and multi shot EPI-FLAIR (MS-EPI-FLAIR). There were negative opinions for the reports of SS-EPI-FLAIR, and it is reported that SS-EPI-FLAIR is able to acquire image quality that is equal to FSE-FLAIR in a shorter time, recently. But there are few reports that utilized contrast of MS-EPI-FLAIR. Our object was to compare HIA detectability in FSE-FLAIR and MS-EPI-FLAIR in equal imaging time, and to evaluate a basic characteristic. We dissolved indigestive dextrin in the aqua destillata and created a 0-46 wt% phantom. The concentration that correspond to T1, T2 value of brain white matter was 42 wt%. We calculated contrast by using signal intensities of each phantom, and compared contrast by signal intensities of each phantom, and compared the contrast of MS-EPI-FLAIR and FSE-FLAIR. In the range of 18-46 wt%, the contrast of MS-EPI-FLAIR was positive value, and that of FSE-FLAIR was 26-42 wt%. By the statistical test, contrast of MS-EPI-FLAIR in 18-26% was significantly different for FSE-FLAIR. In conclusion, it was suggestive that MS-EPI-FLAIR is able to detect HIA equally or sharply.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.6009/jjrt.2015_JSRT_71.12.1180 | DOI Listing |
Nihon Hoshasen Gijutsu Gakkai Zasshi
December 2015
Department of Radiology, Toride Kitasoma Health Medical Center Medical Association Hospital.
In magnetic resonance imaging (MRI) of brain, fast spin echo-fluid attenuated inversion recovery (FSE-FLAIR) is widely used to detect high intensity areas (HIAs). The way of getting FLAIR contrast at high speed includes echo planar imaging-FLAIR (EPI-FLAIR), and in EPI-FLAIR there are single shot EPI-FLAIR (SS-EPI-FLAIR) and multi shot EPI-FLAIR (MS-EPI-FLAIR). There were negative opinions for the reports of SS-EPI-FLAIR, and it is reported that SS-EPI-FLAIR is able to acquire image quality that is equal to FSE-FLAIR in a shorter time, recently.
View Article and Find Full Text PDFActa Radiol
November 2007
Department of Radiology, Akita University School of Medicine, Akita, Japan.
Background: The usefulness of fast fluid-attenuated inversion-recovery (FLAIR) sequences after administration of contrast medium (f-FLAIR (+)) has been shown in depicting brain tumors including metastases and meningeal carcinomatosis. Contrast-enhanced multi-shot echo-planar FLAIR (Ms-EPI-FLAIR (+)), comprising combined sequences of f-FLAIR (+) and Ms-EPI, may provide the advantages of f-FLAIR (+) along with rapid acquisition.
Purpose: To compare Ms-EPI-FLAIR (+) with post-contrast spin-echo T1-weighted imaging (SE-T1WI (+)) in the depiction of brain metastases.