Low level measurement of (60)Co by gamma ray spectrometry using γ-γ coincidence.

Appl Radiat Isot

Laboratoire de Mesure de la Radioactivité dans l'Environnement, IRSN, Orsay, France.

Published: March 2016

This paper presents the latest development of the laboratory to measure the natural and artificial massic activities in environmental samples. The measurement method of coincident emitters by gamma-gamma coincidence using an anti-Compton device and its digital electronics is described. Results obtained with environmental samples are shown. Despite its low efficiency, this method decreases detection limits of (60)Co for certain samples compared to conventional gamma-ray spectrometry due to its very low background.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.apradiso.2015.11.075DOI Listing

Publication Analysis

Top Keywords

environmental samples
8
low level
4
level measurement
4
measurement 60co
4
60co gamma
4
gamma ray
4
ray spectrometry
4
spectrometry γ-γ
4
γ-γ coincidence
4
coincidence paper
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!