Utility of a Newly Designed Film Holder for Premolar Bitewing Radiography.

J Clin Diagn Res

Assistant Professor, Department of Oral and Maxillofacial Radiology, Dental School, Kerman University of Medical Sciences, Kerman, Iran .

Published: November 2015

Introduction: Bitewing radiography is a valuable technique for assessment of proximal caries, alveolar crest and periodontal status. Technical errors during radiography result in erroneous radiographic interpretation, misdiagnosis, possible mistreatment or unnecessary exposure of patient for taking a repeat radiograph.

Aim: In this study, we aimed to evaluate the efficacy of a film holder modified from the conventional one and compared it with that of conventional film holder.

Materials And Methods: Our study population comprised of 70 patients who were referred to the Radiology Department for bilateral premolar bitewing radiographs as requested by their attending clinician. Bitewing radiographs in each patient were taken using the newly designed holder in one side and the conventional holder in the other side. The acceptability of the two holders from the perspectives of the technician and patients was determined using a 0-20 point scale. The frequency of overlap and film positioning errors was calculated for each method.

Results: The conventional holder had greater acceptability among patients compared to the newly designed holder (mean score of 16.59 versus 13.37). From the technicians' point of view, the newly designed holder was superior to the conventional holder (mean score of 17.33 versus 16.44). The frequency of overlap was lower using the newly designed holder (p<0.001) and it allowed more accurate film positioning (p=0.005).

Conclusion: The newly designed holder may facilitate the process of radiography for technicians and may be associated with less frequency of radiographic errors compared to the conventional holder.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4668503PMC
http://dx.doi.org/10.7860/JCDR/2015/13769.6769DOI Listing

Publication Analysis

Top Keywords

newly designed
20
designed holder
16
conventional holder
12
holder
9
film holder
8
premolar bitewing
8
bitewing radiography
8
bitewing radiographs
8
holder side
8
frequency overlap
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!