Background: Alarm fatigue from frequent nonactionable physiologic monitor alarms is frequently named as a threat to patient safety.
Purpose: To critically examine the available literature relevant to alarm fatigue.
Data Sources: Articles published in English, Spanish, or French between January 1980 and April 2015 indexed in PubMed, Cumulative Index to Nursing and Allied Health Literature, Scopus, Cochrane Library, Google Scholar, and ClinicalTrials.gov.
Study Selection: Articles focused on hospital physiologic monitor alarms addressing any of the following: (1) the proportion of alarms that are actionable, (2) the relationship between alarm exposure and nurse response time, and (3) the effectiveness of interventions in reducing alarm frequency.
Data Extraction: We extracted data on setting, collection methods, proportion of alarms determined to be actionable, nurse response time, and associations between interventions and alarm rates.
Data Synthesis: Our search produced 24 observational studies focused on alarm characteristics and response time and 8 studies evaluating interventions. Actionable alarm proportion ranged from <1% to 36% across a range of hospital settings. Two studies showed relationships between high alarm exposure and longer nurse response time. Most intervention studies included multiple components implemented simultaneously. Although studies varied widely, and many had high risk of bias, promising but still unproven interventions include widening alarm parameters, instituting alarm delays, and using disposable electrocardiographic wires or frequently changed electrocardiographic electrodes.
Conclusions: Physiologic monitor alarms are commonly nonactionable, and evidence supporting the concept of alarm fatigue is emerging. Several interventions have the potential to reduce alarms safely, but more rigorously designed studies with attention to possible unintended consequences are needed.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4778561 | PMC |
http://dx.doi.org/10.1002/jhm.2520 | DOI Listing |
Plants (Basel)
January 2025
Department of Interdisciplinary Program in Smart Agriculture, Kangwon National University, Chuncheon 24341, Republic of Korea.
Morphophysiological dormancy (MPD) is considered one of the most primitive dormancy classes among seed plants. While extensive studies have examined the occurrence of endo-β-mannanase in seeds with physiological dormancy (PD) or non-dormancy, little is known about the activity of this enzyme in seeds with MPD. This study aimed to investigate the temporal and spatial patterns of endo-β-mannanase activity during dormancy break and germination.
View Article and Find Full Text PDFNutrients
January 2025
Neonatal Intensive Care Unit, Department of Women's and Children's Health, University Hospital of Padova, 35128 Padova, Italy.
Background: Preterm infants (PIs) are more susceptible to neurodevelopmental impairment compared with term newborns. Adequate postnatal growth has been associated with improved neurocognitive outcomes; therefore, optimization of nutrition may positively impact the neurodevelopment of PIs.
Objective: This study focused on macronutrient parenteral nutrition (PN) intake during the Neonatal Intensive Care Unit stay and their associations with neurodevelopmental outcomes in PIs in the first two years of life.
Sensors (Basel)
January 2025
Department of Biomedical and Robotics Engineering, Incheon National University, Incheon 22012, Republic of Korea.
With the rise of modern healthcare monitoring, heart rate (HR) estimation using remote photoplethysmography (rPPG) has gained attention for its non-contact, continuous tracking capabilities. However, most HR estimation methods rely on stable, fixed sampling intervals, while practical image capture often involves irregular frame rates and missing data, leading to inaccuracies in HR measurements. This study addresses these issues by introducing low-complexity timing correction methods, including linear, cubic, and filter interpolation, to improve HR estimation from rPPG signals under conditions of irregular sampling and data loss.
View Article and Find Full Text PDFSensors (Basel)
January 2025
Department of Biomedical Engineering, Price Faculty of Engineering, University of Manitoba, Winnipeg, MB R3T 2N2, Canada.
Goal: Current methodologies for assessing cerebral compliance using pressure sensor technologies are prone to errors and issues with inter- and intra-observer consistency. RAP, a metric for measuring intracranial compensatory reserve (and therefore compliance), holds promise. It is derived using the moving correlation between intracranial pressure (ICP) and the pulse amplitude of ICP (AMP).
View Article and Find Full Text PDFSensors (Basel)
January 2025
Department of Psychological and Cognitive Sciences, Tsinghua University, Beijing 100084, China.
The objective identification of depression using physiological data has emerged as a significant research focus within the field of psychiatry. The advancement of wearable physiological measurement devices has opened new avenues for the identification of individuals with depression in everyday-life contexts. Compared to other objective measurement methods, wearables offer the potential for continuous, unobtrusive monitoring, which can capture subtle physiological changes indicative of depressive states.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!