COMPUTED TOMOGRAPHY DOSE INDEX MEASUREMENT FOR Hi-ART MEGAVOLTAGE HELICAL CT.

Radiat Prot Dosimetry

Department of Radiation Oncology, Sichuan Cancer Hospital, Chengdu 610041, China.

Published: November 2016

On-line megavoltage computed tomography (MVCT) images are used to verify patient daily set-up in Hi-ART helical TomoTherapy unit. To evaluate the patient dose from MVCT scanning in image guidance, weighted computed tomography (CT) dose index (CTDI) was measured with PTW TM30009 CT pencil chamber in head and body phantoms for slice thicknesses of 2, 4 and 6 mm with different scan lengths. Dose length products (DLPs) were subsequently calculated. The CTDI and DLP were compared with XVI kV CBCT and Brilliance simulator CT for routine clinical protocols. It was shown that CTDI and DLP had close relationship with the slice thickness and the scan length. The dose distribution in the transversal plane was very inhomogeneous due to the attenuation of the couch. Patient dose from MVCT was lower than XVI CBCT for the head scan, while larger for body scan. CTDI, which is measured easily and reproducibly, can be used to assess the patient dose in MVCT. Regular measurement should be performed in QA & QC programmes. Appropriate slice thickness and scan range should be chosen to reduce the patient dose.

Download full-text PDF

Source
http://dx.doi.org/10.1093/rpd/ncv393DOI Listing

Publication Analysis

Top Keywords

patient dose
16
computed tomography
12
dose mvct
12
dose
8
tomography dose
8
ctdi measured
8
ctdi dlp
8
xvi cbct
8
slice thickness
8
thickness scan
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!