Objectives: Compare variability in flow measurements by phase contrast MRI, performed at different locations in the aorta and pulmonary artery (PA) using breath-held (BH) and free-breathing (FB) sequences.

Methods: Fifty-seven patients with valvular heart disease, confirmed by echocardiography, were scanned using BH technique at 3 locations in the ascending aorta (SOV = sinus of Valsalva, STJ = sinotubular junction, ASC = ascending aorta at level of right pulmonary artery) and 2 locations in PA. Single FB measurement was obtained at STJ for aorta. Obtained metrics (SV = stroke volume, FV = forward volume, BV = backward volume, RF = regurgitant fraction) were evaluated separately for patients with aortic regurgitation (AR, n = 31) and mitral regurgitation (n = 26).

Results: No difference was noted between the two measurements in the PA. Significant differences were noted in measured SV at different aortic locations. SV measurements obtained at ASC correlated best with the measurements obtained in the PA. Strongest correlation of AR was measured at the STJ.

Conclusion: Measurements of flow volumes by phase contrast MRI differ depending on slice location. When using stroke volumes to calculate pulmonary to systemic blood flow ratio (Qp/Qs), ASC should be used. For quantifying aortic regurgitation, measurement should be obtained at STJ.

Key Points: • Aortic regurgitation can be accurately measured by MRI. • Aortic regurgitation measurement by MRI varies according to the location where measured. • Aortic regurgitation can also be measured by MRI without breath hold.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00330-015-4120-6DOI Listing

Publication Analysis

Top Keywords

aortic regurgitation
24
• aortic
12
slice location
8
phase contrast
8
contrast mri
8
pulmonary artery
8
regurgitation measurement
8
measured mri
8
regurgitation
7
aortic
6

Similar Publications

We report a 75-year-old female with a history of two heart operations: aortic valve replacement (St. Jude Medical 21 mm) at the age of 44 years for severe rheumatic aortic stenosis and mitral valve replacement (Carbomedics 29 mm) at the age of 51 years for rheumatic mitral regurgitation. Decades later, she presented with exertional dyspnea.

View Article and Find Full Text PDF

Objective: Berry syndrome is a group of rare congenital cardiac malformations including aortopulmonary window (APW), aortic origin of the right pulmonary artery (AORPA), interruption of the aortic arch (IAA), patent ductus arteriosus (PDA) (supplying the descending aorta) and intact ventricular septum. This paper will analyze the clinical data of 7 patients with Berry syndrome who underwent surgical treatment in our institution and discuss the one-stage surgical correction of Berry syndrome in combination with the literature.

Methods: From January 2013 to July 2024, a total of 7 children with Berry syndrome were admitted to the Cardiac Surgery Department of Beijing Children's Hospital.

View Article and Find Full Text PDF

Background: Acute type A aortic dissection (A-AAD) with severe acute aortic regurgitation (AR) and coronary involvement is a potentially fatal condition that causes left ventricular volume overload and catastrophic acute myocardial infarction. We present the successful management of a patient using Impella 5.5 following cardiopulmonary arrest caused by A-AAD with severe acute AR and left main trunk (LMT) obstruction.

View Article and Find Full Text PDF

Background: Right ventricular (RV) function assessment by echocardiography can be challenging due to its complex morphology. Also, increasing use of sedation rather than general anesthesia for transfemoral approach transcatheter aortic valve replacement (TAVR) reduces the need for intraoperative transesophageal echocardiography (TEE). Recent clinical studies have demonstrated the importance of 3-dimensional (3D) echocardiography and a longitudinal strain for RV function assessment.

View Article and Find Full Text PDF

Bicuspid aortic valve (BAV), the most common congenital cardiac anomaly, predisposes individuals to aortic stenosis and regurgitation due to valve degeneration. Abnormal hemodynamics, arterial wall characteristics, and genetic factors contribute to ascending aorta dilatation, potentially leading to severe complications like aortic dissection. Presently, the most recent guidelines propose that individuals with BAV requiring valve replacement due to valve dysfunction should undergo simultaneous replacement of the ascending aorta when the diameter of aortic dilatation exceeds 4.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!