Purpose: To measure sensitivity and stability of the Presage dosimeter in sheet form for various chemical concentrations over a range of clinical photon energies and examine its use for stereotactic body radiation therapy (SBRT) and stereotactic radiosurgery (SRS) QA.
Methods: Presage polymer dosimeters were formulated to investigate and optimize their sensitivity and stability. The dosimeter is composed of clear polyurethane base, leucomalachite green (LMG) reporting dye, and bromoform radical initiator in 0.9-1.0 mm thick sheets. The chemicals are mixed together for 2 min, cast in an aluminum mold, and left to cure at 60 psi for a minimum of two days. Dosimeter response was characterized at energies Co-60, 6 MV, 10 MV flattening-filter free, 15 MV, 50 kVp (mean 19.2 keV), and Ir-192. The dosimeters were scanned by a Microtek Scanmaker i800 at 300 dpi, 2(16) bit depth per color channel. Red component images were analyzed with ImageJ and rit. SBRT QA was done with gamma analysis tolerances of 2% and 2 mm DTA.
Results: The sensitivity of the Presage dosimeter increased with increasing concentration of bromoform. Addition of tin catalyst decreased curing time and had negligible effect on sensitivity. LMG concentration should be at least as high as the bromoform, with ideal concentration being 2% wt. Gamma Knife SRS QA measurements of relative output and profile widths were within 2% of manufacturer's values validated at commissioning, except the 4 mm collimator relative output which was within 3%. The gamma pass rate of Presage with SBRT was 73.7%, compared to 93.1% for EBT2 Gafchromic film.
Conclusions: The Presage dosimeter in sheet form was capable of detecting radiation over all tested photon energies and chemical concentrations. The best sensitivity and photostability of the dosimeter were achieved with 2.5% wt. LMG and 8.2% wt. bromoform. Scanner used should not emit any UV radiation as it will expose the dosimeter, as with the Epson 10000 XL scanner. Presage dosimeter in this form was sensitive enough for use in SRS and SBRT QA. The lower gamma pass rate for Presage compared to Gafchromic film can be attributed to the simple equipment used in the fabrication process, which limited the dosimeter's sensitivity uniformity by agglomeration of air bubbles in the material, nonuniform concentration of chemicals throughout the material, and thickness variations. This demands improvements in mixing tools and molds.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1118/1.4936104 | DOI Listing |
Phys Eng Sci Med
September 2024
School of Physics and Material Studies, Faculty of Applied Sciences, Universiti Teknologi MARA, 40450, Shah Alam, Selangor, Malaysia.
Modern radiotherapy techniques have advanced and become more sophisticated. End-to-end 3D verification of the complex radiotherapy dose distribution in an anthropomorphic phantom can ensure the accuracy of the treatment delivery. The phantoms commonly used for dosimetry are homogeneous solid water phantom which lacks the capability to measure the 3D dose distribution for heterogeneous tissues necessary for advanced radiotherapy techniques.
View Article and Find Full Text PDFMed Phys
October 2023
Department of Radiation Oncology, Seoul National University Hospital, Seoul, Republic of Korea.
Polymers (Basel)
July 2022
School of Physics and Material Studies, Faculty of Applied Sciences, Universiti Teknologi MARA, Shah Alam 40450, Selangor, Malaysia.
Recent advances in radiotherapy technology and techniques have allowed a highly conformal radiation to be delivered to the tumour target inside the body for cancer treatment. A three-dimensional (3D) dosimetry system is required to verify the accuracy of the complex treatment delivery. A 3D dosimeter based on the radiochromic response of a polymer towards ionising radiation has been introduced as the PRESAGE dosimeter.
View Article and Find Full Text PDFPurpose: Two-dimensional (2D) IMRT QA has been widely performed in Radiation Oncology clinic. However, concerns regarding its sensitivity in detecting delivery errors and its clinical meaning have been raised in publications. In this study, a robust methodology of three-dimensional (3D) IMRT QA using fiducial registration and structure-mapping was proposed to acquire organ-specific dose information.
View Article and Find Full Text PDFJ Synchrotron Radiat
November 2020
School of Health and Biomedical Sciences, RMIT University, Plenty Road, Bundoora, Victoria 3083, Australia.
The physical absorbed dose enhancement by the inclusion of gold and bismuth nanoparticles fabricated into water-equivalent PRESAGE dosimeters was investigated. Nanoparticle-loaded water-equivalent PRESAGE dosimeters were irradiated with superficial, synchrotron and megavoltage X-ray beams. The change in optical density of the dosimeters was measured using UV-Vis spectrophotometry pre- and post-irradiation using a wavelength of 630 nm.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!