Introduction: New technology to assist with glenoid placement in shoulder arthroplasty has evolved to include preoperative planning tools and intraoperative guides. These tools provide surgeons with a more complete understanding of glenoid anatomy prior to surgery. However, there have been no studies identifying the information that most influences surgical decision making. Further, there have been few studies that quantify intraoperative identification of scapular landmarks required to execute a preoperative plan. The purpose of this study is to examine the variables that are considered when making a preoperative plan in shoulder arthroplasty.

Methods: The first part of this study was a cadaveric lab in which three surgeons identified the neutral axis in surgical simulation. The second part of the study utilized a preliminary software tool in which surgeons were able to place glenoid implants in a set of CT reconstructions utilizing standard pegged glenoid components. In the third part of the study, surgeons utilized a novel planning software that included the ability to view the 3D reconstructed glenoid in all planes simultaneously and place either standard or augmented glenoid implants. The results of these three studies were compared.

Results: The center of the glenoid identified in the cadaver lab was 1.69 mm ± 1.58 mm anterior and 1.99 mm ± 2.49 mm superior to center. The identified neutral axis was tilted 14.2° ± 9.2° superior to the Friedman axis with 11.8° ± 7.9° of retroversion relative to that axis. Using the novel preoperative planning tool, the surgeons placed implants less than 0.5 mm from the center of the glenoid (AP = -0.07 mm ± 0.42 mm, SI = 0.44 mm ± 0.82 mm) with an average retroversion of less than 1° (-0.96° ± 3.04°).

Conclusion: There was a discernible difference between the neutral axis identified in the cadaveric simulation (aver age of 14.2° superior and 11.8° retroverted) and the implant orientation planned using preoperative software (average of 3.26° superior and 0.96° retroverted). Based on the variability of position and orientation seen cadaverically, it is concluded that additional intraoperative guidance is needed alongside a preoperative plan in order to execute ideal placement of the glenoid component.

Download full-text PDF

Source

Publication Analysis

Top Keywords

preoperative plan
12
neutral axis
12
glenoid
9
shoulder arthroplasty
8
preoperative planning
8
identified neutral
8
tool surgeons
8
glenoid implants
8
center glenoid
8
preoperative
6

Similar Publications

Purpose: To compare the efficacy and safety of skip titanium plates combined with adjacent spinous process suture suspension versus continuous titanium plate fixation in cervical laminoplasty.

Methods: A retrospective analysis of 125 patients (62 men, 63 women, average age 60.9 ± 10.

View Article and Find Full Text PDF

Background: This study aimed to explore the relationship of cervical tumor lesion location (CTLL) with bilateral parametrial involvement (PI) and pelvic lymph node metastasis (LNM).

Methods: The study retrospectively analyzed the clinicopathologic and imaging data of patients with cervical squamous cell carcinoma (SCC) retrieved from multiple centers. According to the CTLL, patients were allocated to three groups: a middle one third group, a unilaterally dominant group, and the entire-region group.

View Article and Find Full Text PDF

Background: Precise and accurate glenoid preparation is important for the success of shoulder arthroplasty. Despite advancements in preoperative planning software and enabling technologies, most surgeons execute the procedure manually. Patient-specific instrumentation (PSI) facilitates accurate glenoid guide pin placement for cannulated reaming; however, few commercially available systems offer depth of reaming control.

View Article and Find Full Text PDF

Mixed reality for preoperative planning and intraoperative assistance of surgical correction of complex congenital heart defects.

J Thorac Cardiovasc Surg

January 2025

Division of Cardiology, The Hospital for Sick Children, Toronto, ON, Canada; Center for Image Guided Innovation and Therapeutic Intervention, The Hospital for Sick Children, Toronto, ON, Canada.

Objectives: Mixed reality (MixR) is an innovative visualization tool that presents virtual elements in a real-world environment, enabling real-time interaction between the user and the combined digital/physical reality. We aimed to explore the feasibility of MixR in enhancing preoperative planning and intraoperative guidance for the correction of various complex congenital heart defects (CHDs).

Methods: Patients underwent cardiac computed tomography or cardiac magnetic resonance and segmentation of digital imaging and communications in medicine (DICOM) images was performed.

View Article and Find Full Text PDF

Is the Coronal Plane Alignment of the Knee (CPAK) Classification Useful to Plan Individualized Total Knee Arthroplasty Surgery for the Spanish Population? A Critical Analysis of the CPAK Classification.

Rev Esp Cir Ortop Traumatol

January 2025

Knee Surgery Unit, iMove Traumatology, Barcelona, Spain; Knee Surgery Unit, Orthopaedic Surgery Department, Hospital Sant Joan de Déu de Manresa - Fundació Althaia, Universitat de Vic, Manresa, Spain.

Introduction: The CPAK classification aims to categorize knee phenotypes. The original study was based on Australian and Belgian population, but significant variation in CPAK distribution exists between different geographic areas. The primary objective is to evaluate knee phenotypes of osteoarthritic Spanish population based on the CPAK system.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!