Introduction And Hypothesis: The objective was to compare the efficacy and safety of an innovative single-incision sling (SIS) with the inside-out transobturator sling (TOT) in the treatment of female stress urinary incontinence (SUI).

Methods: A prospective randomized trial was performed in a tertiary referral urogynecology center from January 2012 to December 2013. The study included women with pure urodynamic SUI. Patients were randomized to either the SIS or the TOT anti-incontinence procedure. Surgery duration, blood loss, and groin pain scores were recorded for each patient. The 1-year follow-up visit included objective and subjective cure parameters, postoperative de novo urgency, complications, and the impact on the patient's life quality.

Results: Of 285 patients assessed for eligibility, a total of 93 patients (32.6 %) were randomized into TOT (n = 48) and SIS groups (n = 45). There were no significant differences in either operating time or blood loss. A statistically significant difference between the two groups was found in pain scores three (5.6 ± 1.8 vs 3.1 ± 2.1, p < 0.001) and 12 h postoperatively (3.8 ± 1.7 vs 2.1 ± 1.7, p < 0.001). After 1 year, there were no significant differences between the TOT and SIS groups in objective cure rates (87.0 % vs 90.9 %; p = 0.399) or patient-reported success rates (91.3 % vs 93.2 %; p = 0.999). Incidence of postoperative de novo urgency did not differ between TOT and SIS patients. Both groups registered a significant improvement in quality of life.

Conclusions: The Ophira SIS procedure has 1-year success rates comparable to standard TOT with significantly less groin pain in the early postoperative period. Both methods were safe and effective in terms of postoperative urgency and life quality improvement.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00192-015-2895-2DOI Listing

Publication Analysis

Top Keywords

prospective randomized
8
single-incision sling
8
stress urinary
8
urinary incontinence
8
1-year follow-up
8
blood loss
8
pain scores
8
randomized comparison
4
comparison transobturator
4
transobturator mid-urethral
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!