Background And Purpose: The metal coverage ratio (MCR) of a flow diverter influences the intra-aneurysmal hemodynamics; a high MCR will occlude an aneurysm early, while a low MCR may delay aneurysm occlusion. The true MCR of a pipeline embolization device (PED) could be lower due to oversize, device deformation, or aneurysm location. In this study deviation of the true MCR from the nominal MCR is assessed and whether their difference affects aneurysm occlusion rate is determined.

Methods: A total of 40 consecutive patients, each of them treated by one PED for their aneurysms at the internal carotid artery (ICA), were retrospectively analyzed. The DynaCT images of these deployed PEDs were used to determine their true dimensions and estimate three MCRs (local, mean, and nominal). These data were compared in two groups of patients who had different aneurysm outcomes at six months.

Results: The difference in the local MCR between two groups is small, but statistically significant (24.5% vs 21.6%, p=05). The local MCR is consistently lower than the nominal MCRs (23.2% vs 30.2%, p<0.001); however, the difference between the mean and local MCRs is small (23.9% vs 23.2%).

Conclusions: An expectation that a PED can achieve a MCR of 30% may not be reasonable. Device oversize and deformation during deployment lower the local MCR by 5-7%. A lowered MCR affects the aneurysm occlusion rate at six months.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4757378PMC
http://dx.doi.org/10.1177/1591019915617315DOI Listing

Publication Analysis

Top Keywords

metal coverage
8
coverage ratio
8
pipeline embolization
8
embolization device
8
mcr
8
aneurysm occlusion
8
true mcr
8
local mcr
8
aneurysm
5
ratio pipeline
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!