Two dedicated software packages -LabSOCS and GESPECOR- for efficiency evaluation in gamma-ray spectrometry, were compared for equivalence. The detection efficiency and the coincidence-summing corrections coefficients were calculated for a specific HPGe detector, for different sample parameters and energies typically encountered in environmental radioactivity measurements. The discrepancy between the results obtained with the two codes were acceptable for most of the applications. Furthermore, the deviations between the values of the standard sources/ reference materials activities from the certificate and the values obtained after Monte Carlo simulation were less than 8% for LabSOCS and 9% for GESPECOR.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.apradiso.2015.11.036 | DOI Listing |