A randomized controlled trial evaluating airway inspection effectiveness during endobronchial ultrasound bronchoscopy.

J Thorac Dis

1 Section of Interventional Pulmonology, Division of Pulmonary and Critical Care, Johns Hopkins Hospital, Johns Hopkins University, Baltimore, MD, USA ; 2 Section of Interventional Pulmonology, Division of Pulmonary and Critical Care, University of North Carolina at Chapel Hill, Chapel Hill, NC, USA ; 3 Department of Biostatistics, the Johns Hopkins Bloomberg School of Public Health, Baltimore, MD, USA.

Published: October 2015

Background: Endobronchial ultrasound transbronchial needle aspiration (EBUS-TBNA) has revolutionized the evaluation of patients with mediastinal and hilar adenopathy. Limitations of conventional endobronchial ultrasound (C-EBUS) bronchoscopes include the inability to perform a complete airway inspection, low definition optics, and limited maneuverability. These limitations require the use of a standard bronchoscope to perform an airway examination prior to the EBUS procedure. Recently, a hybrid endobronchial ultrasound (H-EBUS) bronchoscope with high definition optics and increased maneuverability has been introduced. Our objective was to assess the ability of H-EBUS to perform a full airway inspection and TBNA.

Methods: Patients referred for EBUS-TBNA were prospectively randomized to either form of EBUS from November 2013 to January 2014. The primary outcome was the airway segment visualization in each lobe using an EBUS bronchoscope. Secondary outcomes included the number of bronchoscopes used per procedure, procedure length, diagnostic yield and specimen adequacy.

Results: Sixty-two consecutive patients undergoing EBUS-TBNA were randomized to H-EBUS (n=30) or C-EBUS (n=32). In cases in which EBUS-TBNA was the only procedure performed (n=32), use of a second bronchoscope to perform an adequate airway inspection was significantly higher in C-EBUS compared to H-EBUS (5 vs. 0, P=0.046). There was better segmental visualization achieved in multiple lobes when using H-EBUS (P<0.01). No differences in TBNA sample diagnostic yield, specimen adequacy or procedure time were noted when comparing bronchoscopes (P= NS).

Conclusions: Use of an H-EBUS may improve the ability to perform an adequate airway inspection potentially obviating the need for a conventional bronchoscope.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4635275PMC
http://dx.doi.org/10.3978/j.issn.2072-1439.2015.09.42DOI Listing

Publication Analysis

Top Keywords

airway inspection
16
endobronchial ultrasound
16
definition optics
8
bronchoscope perform
8
airway
6
h-ebus
5
randomized controlled
4
controlled trial
4
trial evaluating
4
evaluating airway
4

Similar Publications

Background: Hemodynamic changes are independent risk factors for long-term patient morbidity and duration of hospital stay. According to the statistics, the three most prevalent complications in the postanaesthesia care unit were airway events, haemodynamic instability, and postoperative nausea and vomiting. Hemodynamic instability in the postanaesthesia care unit can result in serious complications, such as long-term patient morbidity and prolonged hospital stays.

View Article and Find Full Text PDF

Management of the Expected Difficult Airway with Planned One-Lung Ventilation: A Retrospective Analysis of 44 Cases.

Thorac Cardiovasc Surg

November 2024

Department of Anesthesiology, Friedrich-Alexander University Erlangen-Nuremberg, University Hospital Erlangen, Erlangen, Bayern, Germany.

Background:  Difficult airway management is essential in anesthesia practice. Particular challenges are posed to patients who require intraoperative one-lung ventilation. Specific guidelines for these scenarios have been lacking.

View Article and Find Full Text PDF

Background: Common causes of air leakage around an endotracheal tube include insufficient endotracheal tube cuff inflation and damage to the cuff, while damage to the pilot balloon or pilot balloon tubing is relatively rare.

Case Presentation: A 74-year-old female with vertebral osteomyelitis was scheduled for an extreme lateral interbody fusion followed posterior fixation. A fiber-optic light was utilized as part of the surgical illuminator.

View Article and Find Full Text PDF

Continuously monitoring human airway conditions is crucial for timely interventions, especially when airway stents are implanted to alleviate central airway obstruction in lung cancer and other diseases. Mucus conditions, in particular, are important biomarkers for indicating inflammation and stent patency but remain challenging to monitor. Current methods, reliant on computational tomography imaging and bronchoscope inspection, pose risks due to radiation and lack the ability to provide continuous real-time feedback outside of hospitals.

View Article and Find Full Text PDF

Bronchoscopic Diagnosis of Severe Respiratory Infections.

J Clin Med

October 2024

Department of Medicine, Addenbrooke's Hospital, University of Cambridge, Hills Road, Cambridge CB2 0QQ, UK.

The diagnosis of severe respiratory infections in intensive care remains an area of uncertainty and involves a complex balancing of risks and benefits. Due to the frequent colonisation of the lower respiratory tract in mechanically ventilated patients, there is an ever-present possibility of microbiological samples being contaminated by bystander organisms. This, coupled with the frequency of alveolar infiltrates arising from sterile insults, risks over-treatment and antimicrobial-associated harm.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!