Cuff depth and continuous chest auscultation method for determination of tracheal tube insertion depth in nasal intubation: observational study.

J Anesth

Department of Dental Anesthesiology, Field of Oral and Maxillofacial Rehabilitation, Kagoshima University Graduate School of Medical and Dental Sciences, 8-35-1 Sakuragaoka, Kagoshima, 890-8544, Japan.

Published: April 2016

Background: Incorrect endobronchial placement of the tracheal tube can lead to serious complications. Hence, it is necessary to determine the accuracy of tracheal tube positioning. Markers are included on tracheal tubes, in the process of their manufacture, as indicators of approximate intubation depth. In addition, continuous chest auscultation has been used for determining the proper position of the tube. We examined insertion depth using the cuff depth and continuous chest auscultation method (CC method), compared with insertion depth determined by the marker method, to assess the accuracy of these methods.

Methods: After induction of anesthesia, tracheal intubation was performed in each patient. In the CC method, the depth of tube insertion was measured when the cuff had passed through the glottis, and again when breath sounds changed in quality; the depth of tube insertion was determined from these values. In the marker method, the depth of tube insertion was measured and determined when the marker of the tube had reached the glottis, using insertion depth according to the marker as an index.

Results: Insertion depth by the marker method was 26.6 ± 1.2 cm and by the CC method was 28.0 ± 1.2 cm (P < 0.0001). The CC method indicated a significantly greater depth than the marker method.

Conclusion: This study determined the safe range of tracheal tube placement. Tube positions determined by the CC method were about 1 cm deeper than those determined by the marker. This information is important to prevent accidental one-lung ventilation and accidental extubation.

Clinical Trial Registration: UMIN No. UMIN000011375.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00540-015-2110-0DOI Listing

Publication Analysis

Top Keywords

insertion depth
20
tracheal tube
16
tube insertion
16
continuous chest
12
chest auscultation
12
determined marker
12
marker method
12
depth tube
12
depth marker
12
depth
11

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!