Objective: To investigate the feasibility of low-tube-voltage (80 kVp) coronary CT angiography (CCTA) combined with contrast medium (CM) reduction and iterative model reconstruction (IMR) on patients with standard body mass index compared with clinical routine protocol.

Methods: Retrospectively gated helical CCTA scans were acquired using a 256-slice multi-slice CT (Brilliance iCT; Philips Healthcare, Cleveland, OH) on 94 patients with standard body mass index (20-25 kg m(-2)) who were randomly assigned into 2 groups. The scan protocol for Group 1 was 100 kVp and 600 mAs with 70 ml CM at an injection rate of 4.5-5.5 ml s(-1); images were reconstructed by a hybrid iterative reconstruction technique (iDose(4); Philips Healthcare). Group 2 was scanned at 80 kVp and 600 mAs with 35 ml CM at an injection rate of 3.5-4.5 ml s(-1); images were reconstructed with IMR. Objective measurements such as the mean image noise and contrast-to-noise ratio of the two groups were measured on CT images and compared using the paired t-test. In addition, a subjective image quality evaluation was performed by two radiologists who were blinded to the scan protocol, using a 5-point scale [1 (poor) to 5 (excellent)]. The results of the two groups were compared using Mann-Whitney U test.

Results: The iodine delivery rate of Group 2 was 1.0 ± 0.5 gI s(-1) compared with 2.1 ± 0.5 gI s(-1) in Group 1 resulting in a reduction of 52.4%. In addition, an effective radiation dose reduction of 56.4% was achieved in Group 2 (2.4 ± 1.2 mSv) compared with Group 1 (5.5 ± 1.4 mSv). The mean CT attenuation, contrast-to-noise ratio and image quality of all segments in Group 2 were significantly improved compared with those in Group 1 (all, p < 0.01).

Conclusion: The use of IMR along with a low tube voltage (80 kVp) combined with a low CM protocol for CCTA can reduce both radiation and CM dose with improved image quality.

Advances In Knowledge: In this study, we used a novel knowledge-based IMR which remarkably reduced the image noise. We compared the quality of the images obtained when the tube voltage was reduced to 80 kVp and that of those obtained according to the clinical routine protocols to determine whether ultra-low-dose imaging plus IMR is feasible in CCTA scans. We found that a low dose protocol combined with 80 kVp and reduced CM for CCTA can reduce both radiation dose and CM dose with improved image quality by the use of IMR in non-obese patients.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4985220PMC
http://dx.doi.org/10.1259/bjr.20150766DOI Listing

Publication Analysis

Top Keywords

80 kvp coronary
8
coronary angiography
8
combined contrast
8
contrast medium
8
medium reduction
8
reduction iterative
8
iterative model
8
model reconstruction
8
reconstruction imr
8
patients standard
8

Similar Publications

Many nurses and allied professionals (NAPs) lack the skills, knowledge and confidence to engage in conducting and implementing research. This statement describes the importance of NAPs' involvement in clinical research within the context of cardiovascular care. The existing gaps, barriers and enablers to NAPs involvement in research as a potential response to workforce issues in these professions as well as to contribute to excellence in patient care delivery and associated outcomes are identified.

View Article and Find Full Text PDF

Metabolically stable apelin analogs: development and functional role in water balance and cardiovascular function.

Clin Sci (Lond)

January 2025

Center for Interdisciplinary Research in Biology, College de France, Institut National de la Santé et de la Recherche Médicale, Paris, France.

Apelin, a (neuro) vasoactive peptide, plays a prominent role in controlling water balance and cardiovascular functions. Apelin and its receptor co-localize with vasopressin in magnocellular vasopressinergic neurons. Apelin receptors (Apelin-Rs) are also expressed in the collecting ducts of the kidney, where vasopressin type 2 receptors are also present.

View Article and Find Full Text PDF

Importance: A comprehensive lipid panel is recommended by guidelines to evaluate atherosclerotic cardiovascular disease risk, but uptake is low.

Objective: To evaluate whether direct outreach including bulk orders with and without text messaging increases lipid screening rates.

Design, Setting, And Participants: Pragmatic randomized clinical trial conducted from June 6, 2023, to September 6, 2023, at 2 primary care practices at an academic health system among patients aged 20 to 75 years with at least 1 primary care visit in the past 3 years who were overdue for lipid screening.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!