Magnetic resonance imaging investigation of the bone conduction implant - a pilot study at 1.5 Tesla.

Med Devices (Auckl)

Department of Otorhinolaryngology Head and Neck Surgery, Sahlgrenska University Hospital, The Sahlgrenska Academy, University of Gothenburg, Gothenburg, Sweden.

Published: November 2015

Purpose: The objective of this pilot study was to investigate if an active bone conduction implant (BCI) used in an ongoing clinical study withstands magnetic resonance imaging (MRI) of 1.5 Tesla. In particular, the MRI effects on maximum power output (MPO), total harmonic distortion (THD), and demagnetization were investigated. Implant activation and image artifacts were also evaluated.

Methods And Materials: One implant was placed on the head of a test person at the position corresponding to the normal position of an implanted BCI and applied with a static pressure using a bandage and scanned in a 1.5 Tesla MRI camera. Scanning was performed both with and without the implant, in three orthogonal planes, and for one spin-echo and one gradient-echo pulse sequence. Implant functionality was verified in-between the scans using an audio processor programmed to generate a sequence of tones when attached to the implant. Objective verification was also carried out by measuring MPO and THD on a skull simulator as well as retention force, before and after MRI.

Results: It was found that the exposure of 1.5 Tesla MRI only had a minor effect on the MPO, ie, it decreased over all frequencies with an average of 1.1±2.1 dB. The THD remained unchanged above 300 Hz and was increased only at lower frequencies. The retention magnet was demagnetized by 5%. The maximum image artifacts reached a distance of 9 and 10 cm from the implant in the coronal plane for the spin-echo and the gradient-echo sequence, respectively. The test person reported no MRI induced sound from the implant.

Conclusion: This pilot study indicates that the present BCI may withstand 1.5 Tesla MRI with only minor effects on its performance. No MRI induced sound was reported, but the head image was highly distorted near the implant.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4629964PMC
http://dx.doi.org/10.2147/MDER.S90704DOI Listing

Publication Analysis

Top Keywords

tesla mri
16
pilot study
12
implant
9
magnetic resonance
8
resonance imaging
8
bone conduction
8
conduction implant
8
image artifacts
8
test person
8
spin-echo gradient-echo
8

Similar Publications

Sling is Not-Inferior to Brace Immobilization Following Arthroscopic Rotator Cuff Repair: A Randomized Controlled Trial.

Arthroscopy

December 2024

Orthopaedic and Traumatology Unit, Ospedale Regionale di Lugano, EOC, 6900, Lugano, Switzerland; Faculty of Biomedical Sciences, Università della Svizzera Italiana, Via Buffi 13, 6900, Lugano, Switzerland.

Purpose: The aim of this study was to compare the safety and efficacy of immobilizing the upper limb with a brace versus a less-constrained sling in the rehabilitation after arthroscopic rotator cuff repair (ARCR), by documenting clinical and radiological results.

Methods: ARCR was performed in 110 patients (54.9±8.

View Article and Find Full Text PDF

Background And Objectives: Magnetic resonance imaging (MRI) and neurohistopathology are important correlates for evaluation of disease progression in multiple sclerosis (MS). Here we used experimental autoimmune encephalomyelitis (EAE) as an animal model of MS to determine the correlation between clinical EAE severity, MRI and histopathological parameters.

Methods: N = 11 female C57BL/6J mice were immunized with human myelin oligodendrocyte glycoprotein 1-125, while N = 9 remained non-immunized.

View Article and Find Full Text PDF

Cortical lesions impact cognitive decline in multiple sclerosis via volume loss of nonlesional cortex.

Ann Clin Transl Neurol

December 2024

MS Center Amsterdam, Department of Anatomy and Neurosciences, Amsterdam Neuroscience, Vrije Universiteit Amsterdam, Amsterdam, The Netherlands.

Objective: To assess the interrelationship between cortical lesions and cortical thinning and volume loss in people with multiple sclerosis within cortical networks, and how this relates to future cognition.

Methods: In this longitudinal study, 230 people with multiple sclerosis and 60 healthy controls underwent 3 Tesla MRI at baseline and neuropsychological assessment at baseline and 5-year follow-up. Cortical regions (N = 212) were divided into seven functional networks.

View Article and Find Full Text PDF

Purpose: To describe a case in which a right replaced posterior cerebral artery (PCA) was associated with an ipsilateral superior cerebellar artery (SCA) type persistent trigeminal artery (PTA) variant.

Methods: A 53-year-old man who had been diagnosed with chronic dissection of the left vertebral artery (VA) 4 months previously underwent follow-up magnetic resonance (MR) angiography using a 3-Tesla scanner.

Results: MR angiography showed a slightly dilated left VA at the terminal segment without interval change.

View Article and Find Full Text PDF

Background And Purpose: The magnetic resonance images (MRIs) ability of lesion detection in epilepsy is crucial for a diagnosis and surgical outcome. Using automated artificial intelligence (AI)-based tools for measuring cortical thickness and brain volume originally developed for dementia, we aimed to identify whether it could lateralize epilepsy with normal MRIs.

Methods: Non-lesional 3-Tesla MRIs of 428 patients diagnosed with focal epilepsy, based on semiology and electroencephalography findings, were analyzed.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!