Purpose: The purpose of this study was to investigate the role of beam collimator rotation in Volumetric Modulated Arc Therapy (VMAT) for craniospinal irradiation (CSI), and the impact on dose distribution in the beam junctions.
Methods: Six adult patients were selected for the study. Six VMAT plans with different collimator angles were generated for each patient. The patients were treated in supine position with two beam isocenters. The plans were evaluated by analysis of Dose-Volume Histogram (DVHs) data for planning target volume (PTV) and organs at risk (OAR), and conformity index (CI) and homogeneity index (HI) for the target. Dose distributions in the beam junctions were examined carefully and experimentally validated in phantom, with measurement using an ion chamber array and film.
Results: The mean values of HI and CI for the plans with different beam collimator angles were not significantly different. The numbers of segments, monitor units (MUs) and the delivery time of the plans with 45° beam collimator were obviously higher than those in plans with other beam collimator angles. When collimator angle for both sets of beams were set at 0°, there was a 1 mm low dose gap measured in the junction region.
Conclusions: By setting the collimator angle to 45°, only two isocenters were needed for the treatment of a target with the length up to 90 cm. The HI and CI of the plans were almost the same, regardless if the collimator angles were at 0°. The collimator angles for at least one set of beams should be off 0° in order to avoid a dose gap in the beam junction region.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4653929 | PMC |
http://dx.doi.org/10.1186/s13014-015-0544-z | DOI Listing |
Nanophotonics
January 2025
College of Physics & Optoelectronic Engineering, Jinan University, Guangzhou, Guangdong 510632, China.
Grating under auto-collimation configuration with polarization-independent high diffraction efficiency plays an important role in the displacement measurement system, spectral beam combining system and so on. In this paper, we proposed, for the first time, a reflective two-dimensional metal-dielectric grating of which the (-1, -1) order beam is diffracted back along the input light direction, when the incident azimuth angle is 45°. With optimized structure, the (-1, -1) order diffraction efficiencies of transverse electric polarization (TE) and transverse magnetic polarization (TM) are 95.
View Article and Find Full Text PDFNat Commun
January 2025
Department of Electrical and Computer Engineering, National University of Singapore, Singapore, Singapore.
Flat bands have empowered novel phenomena such as robust canalization with strong localization, high-collimation and low-loss propagation. However, the spatial symmetry protection in photonic or acoustic lattices naturally forces flat bands to manifest in pairs aligned at an inherently specific angle, resulting in a fixed bidirectional canalization. Here, we report an acoustic flat-band metasurface, allowing not only unidirectional canalization at all in-plane angles but also robust tunability in band alignment.
View Article and Find Full Text PDFJ Endourol
January 2025
Faculty of Medicine, Department of Urology, Namik Kemal University, Tekirdag, Turkey.
Excessive kidney mobility is an underestimating challenge for surgeons during retrograde intrarenal surgery (RIRS) and extracorporeal shock wave lithotripsy (ESL). There is no technique approved as a gold standard procedure for reducing excessive kidney mobility. The study aimed to uncover predictive factors for determining excessive renal mobility by utilizing clinicodemographic characteristics and noncontrast computed tomography (NCCT) data.
View Article and Find Full Text PDFAnal Chem
January 2025
School of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian, Liaoning 116024, China.
A linear spot-type multipass cell-enhanced fiber-optic photoacoustic gas microprobe is proposed. To further reduce the volume of the gas chamber and enhance the photoacoustic signal, we designed the cross section of the photoacoustic tube as a slit with a height of 10 mm and a width of 1.5 mm.
View Article and Find Full Text PDFMed Phys
January 2025
Department of Radiation Oncology, Inha University Hospital, Incheon, Republic of Korea.
Background: High-dose-rate (HDR) brachytherapy using Iridium-192 as a radiation source is widely employed in cancer treatment to deliver concentrated radiation doses while minimizing normal tissue exposure. In this treatment, the precision with which the sealed radioisotope source is delivered significantly impacts clinical outcomes.
Purpose: This study aims to evaluate the feasibility of a new four-dimensional (4D) in vivo source tracking and treatment verification system for HDR brachytherapy using a patient-specific approach.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!