Titanium-Nitride Coating of Orthopaedic Implants: A Review of the Literature.

Biomed Res Int

Research Centre Linnaeus Institute, Spaarne Hospital, 2134 TM Hoofddorp, Netherlands ; Department of Orthopaedics, Spaarne Hospital, 2134 TM Hoofddorp, Netherlands.

Published: August 2016

Surfaces of medical implants can be enhanced with the favorable properties of titanium-nitride (TiN). In a review of English medical literature, the effects of TiN-coating on orthopaedic implant material in preclinical studies were identified and the influence of these effects on the clinical outcome of TiN-coated orthopaedic implants was explored. The TiN-coating has a positive effect on the biocompatibility and tribological properties of implant surfaces; however, there are several reports of third body wear due to delamination, increased ultrahigh molecular weight polyethylene wear, and cohesive failure of the TiN-coating. This might be due to the coating process. The TiN-coating process should be optimized and standardized for titanium alloy articulating surfaces. The clinical benefit of TiN-coating of CoCrMo knee implant surfaces should be further investigated.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4637053PMC
http://dx.doi.org/10.1155/2015/485975DOI Listing

Publication Analysis

Top Keywords

orthopaedic implants
8
implant surfaces
8
tin-coating
5
titanium-nitride coating
4
coating orthopaedic
4
implants review
4
review literature
4
surfaces
4
literature surfaces
4
surfaces medical
4

Similar Publications

Individualized total knee arthroplasty achieves better functional results than off-the-shelf implants in patients undergoing personalized coronal alignment.

Knee

January 2025

IULS-University Institute for Locomotion and Sports, Pasteur 2 Hospital, University Côte d'Azur, Nice, France; ICARE Team, Côte d'Azur University, Inserm, CNRS, Valrose Institute of Biology, Nice, France. Electronic address:

Background: Several studies have demonstrated the interest in patient-specific custom cutting guides in total knee arthroplasty (TKA), but clinical improvement remains debated. The purpose of this study was to evaluate the functional outcomes (Forgotten Joint Score, FJS) of patients undergoing individualized TKA compared with those receiving off-the-shelf (OTS) implants, both using patient-specific cutting guides with personalized alignment over a minimum follow up period of 12 months. We hypothesized that individualized TKA demonstrates significantly better functional outcomes than OTS TKA (FJS and percentage of patients reaching the minimum clinically important difference).

View Article and Find Full Text PDF

Purpose: Vertebral body tethering (VBT) is a non-fusion surgical option for adolescent idiopathic scoliosis (AIS) that requires a postoperative (PO) chest tube. This study evaluates whether 48 h of PO TXA reduces chest tube (CT) drainage and retention compared to 24 h of TXA following VBT for AIS.

Methods: Consecutively treated patients with a diagnosis of AIS who underwent VBT were assessed.

View Article and Find Full Text PDF

The best treatment method for reverse obliquity intertrochanteric fractures (ROIFs) is still under debate. Our team designed the modified proximal femoral nail (MPFN) specially for treating such fractures. The objective of this research was to introduce the MPFN device and compare the biomechanical properties with Proximal Femoral Nail Antirotation (PFNA) and InterTAN nail via finite element modelling.

View Article and Find Full Text PDF

Introduction: As the number of revision total knee arthroplasties (rTKA) continues to rise, there is increasing interest in the use of contemporary rotating hinge prostheses. These devices often incorporate porous cones to fill bone defects and enhance long-term fixation. This study evaluated the clinical and functional outcomes and survivorship in rTKA patients utilizing a rotating hinge prosthesis with flexible titanium (FT) cones, porous tantalum (PT) cones, or no cones.

View Article and Find Full Text PDF

A free-floating disc shaped polycarbonate-urethane ultra-high molecular weight polyethylene fiber reinforced medial compartment implant is designed for symptomatic postmedial meniscectomy syndrome. Because it is not sutured into place, an intact 2mm meniscus rim with intact anterior and posterior meniscal horns are required. In a recent 24-month follow-up study, only 64% of the original implants were retained.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!