Study Objectives: Pressure-relief features are aimed at improving the patient's comfort during continuous positive airway pressure (CPAP) treatment for obstructive sleep apnea. The objective of this study was to determine the effect of these therapy features on fixed CPAP and autotitrating CPAP (APAP) treatment efficacy.
Methods: Seven pressure-relief features applied by three CPAP devices were included in our study (Remstar Auto: C-Flex 3, C-Flex+ 3, A-Flex 3, P-Flex; AirSense 10: EPR 3; Prisma 20A: SoftPAP 2 and 3). In fixed CPAP, the devices were subjected to a 10-min bench-simulated obstructive apnea sequence (initial apnea-hypopnea index, AHI = 60/h) with and without pressure-relief features. In APAP, the sequence was lengthened to 4.2 h (initial AHI = 58.6/h). The residual AHI and mean/median pressure were compared with and without pressure-relief features.
Results: Compared to conventional CPAP, where pressure was adjusted to be just sufficient to control the simulated obstructive events, C-Flex+ 3, P-Flex, and EPR 3 failed to normalize the breathing flow and did not reduce the AHI. The mean pressures with the three features, respectively, were 1.8, 2.6, and 2.6 cmH2O lower than the conventional CPAP. Compared to conventional APAP, similar levels of control were observed with pressure-relief features, apart from P-Flex where the delivered mean pressure was lower and residual AHI greater. The device-reported mean/median pressures in APAP with A-Flex 3, P-Flex, EPR 3, and SoftPAP 3 were higher than that measured on the bench.
Conclusions: Pressure-relief features may attenuate CPAP efficacy if not adjusted for at the time of their introduction. In clinical practice, efficacy can be ensured by increasing the therapeutic pressure delivered by fixed CPAP or by enabling the pressure-relief features prior to initial pressure titration. Device-reported pressures in APAP devices with pressure relief activated may overstate delivered pressures.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4773618 | PMC |
http://dx.doi.org/10.5664/jcsm.5590 | DOI Listing |
A A Pract
July 2024
From the Department of Anesthesia and Perioperative Care, University of California, San Francisco, San Francisco, California.
The anesthesia gas scavenging system (AGSS) removes waste gases from the anesthesia machine. Within the AGSS, safety features prevent excessive pressures from affecting ventilation. Although the literature contains reports describing failures of the AGSS, we found no reports of positive-pressure relief valve (PPRV) malfunctions.
View Article and Find Full Text PDFACS Omega
December 2023
The University of Texas Permian Basin, Odessa, Texas 79762, United States.
In the oil and gas industry, gas flaring through flare stacks plays a critical role in safely and efficiently releasing and burning gases and other materials from pressure-relief and vapor-depressurizing systems. Gas flaring is a contentious and formidable environmental challenge, both regionally and globally. Its adverse impacts encompass pollution, contribution to global warming, and substantial economic losses.
View Article and Find Full Text PDFDisabil Rehabil Assist Technol
October 2024
Integrated Design and Management, Massachusetts Institute of Technology, Cambridge, MA, USA.
Pressure reliefs (PRs) are self-care practices essential for individuals with spinal cord injury (SCI) to prevent life-threatening pressure injuries (PIs). Despite the benefits, individuals often do not do these exercises at home, leading to increased patient morbidity and mortality. To examine how digital technology could improve this population's adherence to PR exercises, we conducted a technology probe study with five individuals with SCI over ten consecutive business days.
View Article and Find Full Text PDFObjective: Obstructive sleep apnea (OSA) is the most common type of sleep-disordered breathing. This article describes advances in the diagnosis, testing, treatment, and monitoring of OSA.
Latest Developments: Home sleep apnea testing and in-laboratory polysomnography are the most commonly used diagnostic tools in the identification and monitoring of OSA, but new methods for diagnosis and at-home monitoring of treatment response are being developed and validated.
J Biomech
May 2023
Faculty of Sports Science, Ningbo University, Ningbo, China. Electronic address:
A carbon-fiber plate (CFP) embedded into running shoes is a commonly applied method to improve running economy, but little is known in regard the effects of CFP design features on internal foot mechanics. This study aimed to explore how systematic changes in CFP geometrical variations (i.e.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!