Background: The teeth with undiagnosed vertical root fractures (VRFs) are likely to receive endodontic treatment or retreatment, leading to frustration and inappropriate endodontic therapies. Moreover, many cases of VRFs cannot be diagnosed definitively until the extraction of tooth.

Objectives: This study aimed to assess the use of different voxel resolutions of two different cone beam computerized tomography (CBCT) units in the detection VRFs in vitro.

Materials And Methods: The study material comprised 74 extracted human mandibular single rooted premolar teeth without root fractures that had not undergone any root-canal treatment. Images were obtained by two different CBCT units. Four image sets were obtained as follows: 1) 3D Accuitomo 170, 4 × 4 cm field of view (FOV) (0.080 mm(3)); 2) 3D Accuitomo 170. 6 × 6 cm FOV (0.125 mm(3)); 3) NewTom 3G, 6˝ (0.16 mm(3)) and 4) NewTom 3G, 9˝ FOV (0.25 mm(3)). Kappa coefficients were calculated to assess both intra- and inter-observer agreements for each image set.

Results: No significant differences were found among observers or voxel sizes, with high average Z (Az) results being reported for all groups. Both intra- and inter-observer agreement values were relatively better for 3D Accuitomo 170 images than the images from NewTom 3G. The highest Az and kappa values were obtained with 3D Accuitomo 170, 4 × 4 cm FOV (0.080 mm(3)) images.

Conclusion: No significant differences were found among observers or voxel sizes, with high Az results reported for all groups.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4632561PMC
http://dx.doi.org/10.5812/iranjradiol.18290DOI Listing

Publication Analysis

Top Keywords

accuitomo 170
16
voxel resolutions
8
vertical root
8
root fractures
8
cbct units
8
fov 0080
8
0080 mm3
8
170 fov
8
mm3 newtom
8
intra- inter-observer
8

Similar Publications

Purpose: This study investigated the impact of the field of view (FOV), voxel size, and exposure parameters on the contrast-to-noise ratio (CNR) in cone-beam computed tomography (CBCT).

Materials And Methods: A SedentexCT phantom was scanned using 3D Accuitomo 170 across 3 FOVs (40×40 mm, 60×60 mm, 80×80 mm). Each FOV had 4 settings for kVp and 3 for mA.

View Article and Find Full Text PDF

Optimization of orofacial cleft imaging protocols using device-specific low-dose cone-beam computed tomography.

J Oral Rehabil

September 2024

OMFS IMPATH Research Group, Department of Imaging & Pathology, Faculty of Medicine, KU Leuven & Oral and Maxillofacial Surgery, University Hospitals Leuven, Leuven, Belgium.

Objective: The aim of this study was to present optimized device-specific low-dose cone-beam computed tomography (CBCT) protocols with sufficient image quality for pre-surgical diagnostics and three-dimensional (3D) modelling of cleft defects.

Methods: Six paediatric skulls were acquired, and an artificial bony cleft was created. A high-resolution CBCT scan acted as a reference standard (Accuitomo 170, Morita, Kyoto, Japan) for comparing eight low-dose protocols of Newtom VGi-evo (QR Verona, Cefla, Verona, Italy), which included Eco and Regular protocols with different field of views (FOVs).

View Article and Find Full Text PDF
Article Synopsis
  • The study investigates the anatomical differences of tooth roots and canals in a Saudi Arabian population using advanced imaging techniques like cone-beam computed tomography (CBCT) and micro-computed tomography (micro-CT).
  • A systematic review of 47 studies revealed that the majority utilized CBCT for analyzing a total of 47,612 tooth samples, focusing on both maxillary and mandibular teeth, while micro-CT studies were more limited.
  • The quality of the included studies was assessed for bias risk, and several classifications (predominantly Vertucci's) were applied to categorize the root canal morphology found in the samples.
View Article and Find Full Text PDF

Assessment of the root filling volume in C-shaped root canal on cone-beam CT and micro-CT in relation to nano-CT.

Clin Oral Investig

November 2023

Department of Restorative Dentistry, School of Dentistry of Ribeirão Preto, University of São Paulo, Av. do Café-Subsetor Oeste-11 (N-11), Ribeirão Preto, São Paulo, 14040-904, Brazil.

Aim: To assess the observed volume of filled C-shaped root canals from different CBCT and micro-CT having nano-CT as a reference.

Materials And Methods: Twelve extracted mandibular molars with C-shaped root canals were endodontically treated using Reciproc Blue R25 (VDW GMBH, Munich, Germany) in a reciprocating system and filled with gutta-percha (Dentsply Maillefer) and AH Plus sealer (Dentsply DeTrey, Konstanz, Germany). CBCT scans were performed using 3 units-3D Accuitomo 170 (J.

View Article and Find Full Text PDF

Photon-counting computed tomography (PCCT) is an innovative technological advancement in relation to x-ray detectors which offers ultra-high-resolution images. The current study aimed to evaluate the visualization ability of PCCT compared to cone-beam computed tomographic (CBCT) devices for challenging endodontic diagnostic tasks. A reference image of an anthropomorphic phantom was acquired using an industrial micro-CT device.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!