Introduction: Available assays for quantitation of the Alzheimer's disease (AD) biomarker amyloid-beta 1-42 (Aβ [1-42]) in cerebrospinal fluid demonstrate significant variability and lack of standardization to reference measurement procedures (RMPs). We report analytical performance data for the novel Elecsys β-amyloid (1-42) assay (Roche Diagnostics).

Methods: Lot-to-lot comparability was tested using method comparison. Performance parameters were measured according to Clinical & Laboratory Standards Institute (CLSI) guidelines. The assay was standardized to a Joint Committee for Traceability in Laboratory Medicine (JCTLM) approved RMP.

Results: Limit of quantitation was <11.28 pg/mL, and the assay was linear throughout the measuring range (200-1700 pg/mL). Excellent lot-to-lot comparability was observed (correlation coefficients [Pearson's r] >0.995; bias in medical decision area <2%). Repeatability coefficients of variation (CVs) were 1.0%-1.6%, intermediate CVs were 1.9%-4.0%, and intermodule CVs were 1.1%-3.9%. Estimated total reproducibility was 2.0%-5.1%. Correlation with the RMP was good (Pearson's r, 0.93).

Discussion: The Elecsys β-amyloid (1-42) assay has high analytical performance that may improve biomarker-based AD diagnosis.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jalz.2015.09.009DOI Listing

Publication Analysis

Top Keywords

cerebrospinal fluid
8
technical performance
4
performance novel
4
novel fully
4
fully automated
4
automated electrochemiluminescence
4
electrochemiluminescence immunoassay
4
immunoassay quantitation
4
quantitation of β-amyloid
4
of β-amyloid 1-42
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!