The aim of this in vitro study was to compare the actual torque of 2 manual wrench systems to their stated (target) torque. New spring- (Nobel Biocare USA, LLC) and friction-style (Zimmer Dental, Inc.) manual dental torque wrenches, as well as spring torque wrenches that had undergone sterilization and clinical use, were tested. A calibrated torque gauge was used to compare actual torque to target torque values of 15 and 35 N/cm. Data were statistically analyzed via mixed-effects regression model with Bonferroni correction. At a target torque of 15 N/cm, the mean torque of new spring wrenches (13.97 N/cm; SE, 0.07 N/cm) was significantly different from that of used spring wrenches (14.94 N/cm; SE, 0.06 N/cm; P < 0.0001). However, the mean torques of new spring and new friction wrenches (14.10 N/cm; SE, 0.07 N/cm; P = 0.21) were not significantly different. For torque measurements calibrated at 35 N/cm, the mean torque of new spring wrenches (35.29 N/cm; SE, 0.10 N/cm) was significantly different (P < 0.0001) from the means of new friction wrenches (36.20 N/cm; SE, 0.08 N/cm) and used spring wrenches (36.45 N/cm; SE, 0.08 N/cm). Discrepancies in torque could impact the clinical success of screw-retained dental implants. It is recommended that torque wrenches be checked regularly to ensure that they are performing to target values.
Download full-text PDF |
Source |
---|
J Shoulder Elbow Surg
December 2024
Orthopedic Biomechanics Laboratory, Congress Medical Foundation, Pasadena, CA, USA. Electronic address:
Background: The restriction of active internal rotation (IR) after reverse shoulder arthroplasty (RSA) poses a challenging problem for reconstructive shoulder surgeons, particularly in patients suffering from massive rotator cuff tears (mRCT) with subscapularis (SSC) deficiency. This study aims to evaluate the biomechanical effectiveness of different tendon transfer techniques following medialized glenoid and lateralized humerus RSA in improving internal rotation (IR) strength.
Methods: Eight cadaveric shoulder specimens were evaluated using a custom shoulder testing system designed to simulate loading conditions typical of mRCT with SSC insufficiency.
J Prosthet Dent
November 2024
Associate Professor, State Key Laboratory of Military Stomatology & National Clinical Research Center for Oral Diseases & Shaanxi Engineering Research Center for Dental Materials and Advanced Manufacture, Department of Implant Dentistry, School of Stomatology, The Air Force Medical University, Xi'an, PR China. Electronic address:
Statement Of Problem: Three months after implant placement, difficulty was encountered in unscrewing the healing abutment within the normal torque range (<35 Ncm), resulting in the removal of the implant. Despite all attempts, the implant could not then be separated from the healing abutment. Retrospective analysis of intraoperative photographs revealed the presence of blood or bone debris within the implant before the healing abutment was placed; this was considered to be the likely cause of this phenomenon.
View Article and Find Full Text PDFJ Prosthet Dent
December 2024
Professor, School of Dentistry, University of Ribeirão Preto (UNAERP), Ribeirão Preto, Brazil. Electronic address:
Statement Of Problem: The biomechanical stability of the implant-prosthesis assembly and its maintenance under function is a determining factor in the success of implant prosthesis rehabilitation, but studies of different angular tapers are lacking.
Purpose: The purpose of this in vitro study was to evaluate the axial displacement and torque loss of prosthetic abutments with Morse-type connections of different angular tapers after thermomechanical cycling by using microcomputed tomography (µCT) and a digital torque wrench.
Material And Methods: Eighteen Ø3.
Asian Spine J
October 2024
Faculty of Engineering, Yamaguchi University, Ube, Japan.
Study Design: Biomechanical study.
Purpose: To investigate the mechanical characteristics of bone models created from medical images.
Overview Of Literature: Recent advancements in three-dimensional (3D) printing technology have affected its application in surgery.
BMC Musculoskelet Disord
June 2024
Orthopedic Department, 960 Hospital of People's Liberation Army, NO.25 Shifan Road, Jinan, Shandong, 250031, China.
Purpose: To analyze the characteristics of PEEK rods retrieved in vivo, specifically their wear and deformation, biodegradability, histocompatibility, and mechanical properties.
Method: Six PEEK rods were retrieved from revision surgeries along with periprosthetic tissue. The retrieved PEEK rods were evaluated for surface damage and internal changes using Micro-CT, while light and electron microscopy were utilized to determine any histological changes in periprosthetic tissues.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!