Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: The FloTrac/Vigileo™ system does not thoroughly reflect variable arterial tones, due to a lack of external calibration. The ability of this system to measure stroke volume and track its changes after fluid administration has not been fully evaluated in patients with the high systemic vascular resistance that can develop during laparoscopic surgery.
Methods: In 42 patients undergoing laparoscopic prostatectomy, the stroke volume derived by the third-generation FloTrac/Vigileo™ system (SV-Vigileo), the stroke volume measured using transesophageal echocardiography (SV-TEE) as a reference method, and total systemic vascular resistance were evaluated before and after 500 ml fluid administration during pneumoperitoneum combined with the Trendelenburg position.
Results: Total systemic vascular resistance was 2159.4 ± 523.5 dyn·s/cm5 before fluid administration. The SV-Vigileo was significantly higher than the SV-TEE both before (68.8 ± 15.9 vs. 57.0 ± 11.0 ml, P < 0.001) and after (73.0 ± 14.8 vs. 64.9 ± 12.2 ml, P = 0.003) fluid administration. During pneumoperitoneum combined with the Trendelenburg position, Bland-Altman analysis for repeated measures showed a 53.8% of percentage error between the SV-Vigileo and the SV-TEE. Four-quadrant plot (69.2% of a concordance rate) and polar plot analysis (20.6° of a mean polar angle, 16.4° of the SD of a polar angle, and ±51.5° of a radial sector containing 95% of the data points) did not indicate a good trending ability of the FloTrac/Vigileo™ system.
Conclusions: The third-generation FloTrac/Vigileo™ system may not be useful in patients undergoing laparoscopic surgery, based on unreliable performance in measuring the stroke volume and in tracking changes in the stroke volume after fluid administration during pneumoperitoneum combined with the Trendelenburg position.