Objective: To detect platelet anti-HPA-1a and -1b antibodies using recombinant GPIIIa fragments coupled to Luminex beads.

Methods: The sensitivity of 2 techniques, monoclonal antibody specific immobilization of platelet antigen (MAIPA) and Luminex bead assay, was compared using 12 twofold-serial dilutions (from neat to 1 in 2048) of an anti-HPA-1a WHO international standard. The specificity of Luminex assay to identify anti-HPA-1a and -1b antibodies was assessed using 8 negative or positive controls and 36 blinded samples provided by WHO Platelet Workshop.

Results: The sensitivity of MAIPA and Luminex bead assay to detect anti-HPA-1a was dilution 1/64 (i.e. 1.56 IU/ml) and far more than dilution 1/2048 (i.e. 0.049 IU/mL), respectively. The Luminex bead assay could specifically identify negative and positive controls of anti-HPA-1a and -1b. All results of 36 blinded samples by Luminex assay were accordant to reference results except one sample which contained high concentration antithetical antibody and resulted in false positive of anti-HPA-1b. Cross-reactivity was also not observed with the samples containing HLA, ABO or other platelet antibodies.

Conclusion: The Luminex beads coupled with recombinant GPIIIa fragments can be used to detect HPA-1 system antibodies with sufficient sensitivity and specificity, that is suitable for the detection of platelet alloantibodies in clinical alloimmune thrombocytopenia.

Download full-text PDF

Source
http://dx.doi.org/10.7534/j.issn.1009-2137.2015.05.031DOI Listing

Publication Analysis

Top Keywords

recombinant gpiiia
12
gpiiia fragments
12
anti-hpa-1a -1b
12
luminex bead
12
bead assay
12
fragments coupled
8
luminex
8
coupled luminex
8
-1b antibodies
8
maipa luminex
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!