Purpose: To develop and establish a comprehensive gating commissioning and quality assurance procedure in compliance with TG-142.

Methods: Eight Varian TrueBeam Linacs were used for this study. Gating commissioning included an end-to-end test and baseline establishment. The end-to-end test was performed using a CIRS dynamic thoracic phantom with a moving cylinder inside the lung, which was used for carrying both optically simulated luminescence detectors (OSLDs) and Gafchromic EBT2 films while the target is moving, for a point dose check and 2D profile check. In addition, baselines were established for beam-on temporal delay and calibration of the surrogate, for both megavoltage (MV) and kilovoltage (kV) beams. A motion simulation device (MotionSim) was used to provide periodic motion on a platform, in synchronizing with a surrogate motion. The overall accuracy and uncertainties were analyzed and compared.

Results: The OSLD readings were within 5% compared to the planned dose (within measurement uncertainty) for both phase and amplitude gated deliveries. Film results showed less than 3% agreement to the predicted dose with a standard sinusoid motion. The gate-on temporal accuracy was averaged at 139±10 ms for MV beams and 92±11 ms for kV beams. The temporal delay of the surrogate motion depends on the motion speed and was averaged at 54.6±3.1 ms for slow, 24.9±2.9 ms for intermediate, and 23.0±20.1 ms for fast speed.

Conclusions: A comprehensive gating commissioning procedure was introduced for verifying the output accuracy and establishing the temporal accuracy baselines with respiratory gating. The baselines are needed for routine quality assurance tests, as suggested by TG-142.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.4932363DOI Listing

Publication Analysis

Top Keywords

quality assurance
12
gating commissioning
12
assurance tests
8
respiratory gating
8
comprehensive gating
8
end-to-end test
8
temporal delay
8
surrogate motion
8
temporal accuracy
8
motion
6

Similar Publications

Validity of one-time phantomless patient-specific quality assurance in proton therapy with regard to the reproducibility of beam delivery.

Med Phys

January 2025

OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Helmholtz-Zentrum Dresden - Rossendorf, Dresden, Germany.

Background: Patient-specific quality assurance (PSQA) is a crucial yet resource-intensive task in proton therapy, requiring special equipment, expertise and additional beam time. Machine delivery log files contain information about energy, position and monitor units (MU) of all delivered spots, allowing a reconstruction of the applied dose. This raises the prospect of phantomless, log file-based QA (LFQA) as an automated replacement of current phantom-based solutions, provided that such an approach guarantees a comparable level of safety.

View Article and Find Full Text PDF

Background: Online adaptive radiotherapy (OART) and rapid quality assurance (QA) are essential for effective heavy ion therapy (HIT). However, there is a shortage of deep learning (DL) models and workflows for predicting Monte Carlo (MC) doses in such treatments.

Purpose: This study seeks to address this gap by developing a DL model for independent MC dose (MCDose) prediction, aiming to facilitate OART and rapid QA implementation for HIT.

View Article and Find Full Text PDF

Introduction: The global prevalence of antimicrobial resistance transcends geographical and economic boundaries, affecting populations worldwide. Excessive and incorrect use of antibiotics encourages antimicrobial resistance which leads to complex treatment strategies for infectious diseases and possible failure of treatment. The incorrect and unnecessary prescribing of antibiotics places a burden on healthcare costs and thus, antimicrobial resistance is evident globally as a major public health concern.

View Article and Find Full Text PDF

Host response to environmental exposures such as pathogens and chemicals can include modifications to the epigenome and transcriptome. Improved signature discovery, including the identification of the agent and timing of exposure, has been enabled by advancements in assaying techniques to detect RNA expression, DNA base modifications, histone modifications, and chromatin accessibility. The interrogation of the epigenome and transcriptome cascade requires analyzing disparate datasets from multiple assay types, often at single-cell resolution, derived from the same biospecimen.

View Article and Find Full Text PDF

Background: To date there has been no evaluation of further training and working conditions in medical specialist training in psychiatry and psychotherapy in Germany. In order to counteract the lack of young talent and ensure long-term care for patients, quality assurance is essential.

Aim Of The Work: The evaluation aims to identify strengths and weaknesses of the further training and to derive measures for improvement.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!