Multi-layer Flow-modulating Stents for Thoraco-abdominal and Peri-renal Aneurysms: The UK Pilot Study.

Eur J Vasc Endovasc Surg

Department of Academic Surgery, Institute of Cardiovascular Sciences, University of Manchester, Education and Research Centre, University Hospital South Manchester, Manchester, UK; Department of Vascular and Endovascular Surgery, University Hospital South Manchester, Manchester, UK.

Published: February 2016

Objective: There remains a population of patients with aortic aneurysms that cannot be treated by conventional endovascular means. Multi-layer flow modulating stents (MFMS) are a novel approach for the treatment of aortic aneurysm; this study reports outcomes of a UK pilot study of first-generation MFMS in thoraco-abdominal (TAAA) and perirenal aneurysms (PAA) in patients who were also unfit for open surgery.

Methods: Patients with TAAA and PAA unfit for open surgery and with no conventional options for endovascular repair were recruited. Follow-up included CTA at 1, 3, 6, and 12 months, then annually. Outcome measures included 30 day mortality, growth-free survival, branch vessel patency, complications, re-intervention, and maximal aortic diameter.

Results: MFMS were implanted in 14 patients (6 PAA, 8 TAAA) between October 2011 and March 2014 with one (7%) 30 day death and 11 (79%) surviving to 12 months. The median aneurysm growth was 9 mm in the first 12 months following implantation. On mean follow-up of 22.8 months, seven (50%) patients had died including one confirmed rupture. AAA diameter remained stable in only two of the surviving patients. Fifty of 51 covered aortic branches remained patent with no embolic episodes or symptoms of ischaemia in any patient. MFMS dislocation occurred in four patients, leading to re-intervention in two. A total of six re-interventions were performed in five patients (35%) with one post-re-intervention death.

Conclusion: These first-generation MFMS were unstable and dislocated frequently. It is uncertain whether MFMS implantation influenced the natural history of these aneurysms as none decreased in size, but two remain stable after a mean of 22.8 months. Although side branch patency was maintained, our results do not support the continued use of these first-generation devices. Further development is needed if this technology is to have a role in treatment of aortic aneurysm.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ejvs.2015.09.014DOI Listing

Publication Analysis

Top Keywords

pilot study
8
patients
8
treatment aortic
8
aortic aneurysm
8
first-generation mfms
8
unfit open
8
228 months
8
mfms
6
aortic
5
months
5

Similar Publications

Newborn screening in Brazil: realities and challenges.

J Community Genet

January 2025

Laboratório de Hormônios e Genética Molecular-LIM42, Disc de Endocrinologia do Hospital das Clínicas, da Faculdade de Medicina da Universidade de São Paulo, São Paulo, Brazil.

Neonatal screening is a critical public health initiative introduced worldwide to detect severe congenital disorders early in life. This study provides an updated overview of public neonatal screening in Brazil, highlighting the transition from initial pilot projects to a robust national program. Through the analysis of recent data up to 2023, we discuss the coverage and regional efficiencies of the program.

View Article and Find Full Text PDF

Sleep disorders in individuals with dentofacial deformities: a preliminary study on the relationship with sex and serotonin levels.

Oral Maxillofac Surg

January 2025

Programa de Pós-graduação em Odontologia, Escola de Ciências da Saúde e da Vida, Pontifícia Universidade Católica do Rio Grande do Sul, Porto Alegre, RS, Brazil.

Purpose: It has been suggested that dentofacial deformities (DFD) can impair sleep quality. This pilot study aimed at evaluating sleep disorders in individuals with DFD before orthognathic surgery, correlating the clinical findings with salivary biomarker levels.

Methods: This cross-sectional study enrolled ten males and ten females with DFD diagnoses under orthodontic treatment preceding orthognathic surgery.

View Article and Find Full Text PDF

The underlying mechanisms explaining the differential course of SARS-CoV-2 infection and the potential clinical consequences after COVID-19 resolution have not been fully elucidated. As a dysregulated mitochondrial activity could impair the immune response, we explored long-lasting changes in mitochondrial functionality, circulating cytokine levels, and metabolomic profiles of infected individuals after symptoms resolution, to evaluate whether a complete recovery could be achieved. Results of this pilot study evidenced that different parameters of aerobic respiration in lymphocytes of individuals recuperated from a severe course lagged behind those shown upon mild COVID-19 recovery, in basal conditions and after simulated reinfection, and they also showed altered glycolytic capacity.

View Article and Find Full Text PDF

Accuracy and feasibility of continuous glucose monitoring system in pancreatectomy patients.

Langenbecks Arch Surg

January 2025

Department of Surgery, Division of HBP Surgery & Liver Transplantation, Korea University College of Medicine, Seoul, Korea.

Purpose: Pancreatectomy patients often experience challenging fluctuations in blood glucose levels; therefore, they require a reliable monitoring system. This study aimed to determine the accuracy and acceptability of a continuous glucose monitoring (CGM) system compared with the intermittent capillary glucose test in patients who have undergone pancreatectomy.

Methods: Thirty non-diabetic pancreatectomy patients participated.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!