Objective: To design new apparatus and develop relative surgical procedures for all-arthroscopic tibial inlay PCL reconstruction.

Methods: The new apparatus consisting of three main components for total arthroscopic tibial inlay were designed and manufactured, and the relative surgical procedures were experimented on 6 cadaveric knees.

Results: The apparatus for tibial sloting including three components was designed and manufactured successfully, which could be used easily. The surgical procedures were proved to be feasible, which could be performed within a short time (mean, 78.5 min). The graft fixation was proved to be firm enough by mechanical tests. Image examination (X ray, CT and MRI) confirmed that the location of implants was appropriate.

Conclusion: The newly designed apparatus and surgical procedures for all-arthroscopic tibial inlay seems feasible and useful.

Download full-text PDF

Source

Publication Analysis

Top Keywords

tibial inlay
16
surgical procedures
16
relative surgical
8
procedures all-arthroscopic
8
all-arthroscopic tibial
8
designed manufactured
8
tibial
5
novel all-arthroscopic
4
all-arthroscopic pcl
4
pcl tibial
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!