Background: Maintaining the cuff pressure of endotracheal tubes (ETTs) within 20-30 cmH2O is a standard practice. The aim of the study was to evaluate the effectiveness of standard practice in maintaining cuff pressure within the target range.

Methods: This was a prospective observational study conducted in a tertiary-care intensive care unit, in which respiratory therapists (RTs) measured the cuff pressure 6 hourly by a handheld manometer. In this study, a research RT checked cuff pressure 2-4 h after the clinical RT measurement. Percentages of patients with cuff pressure levels above and below the target range were calculated. We identified predictors of low-cuff pressure.

Results: We analyzed 2120 cuff-pressure measurements. The mean cuff pressure was 27 ± 2 cmH2O by the clinical RT and 21 ± 5 cmH2O by the research RT (p < 0.0001). The clinical RT documented that 98.0 % of cuff pressures were within the normal range. The research RT found the cuff pressures to be within the normal range in only 41.5 %, below the range in 53 % and above the range in 5.5 %. Low cuff pressure was found more common with lower ETT size (OR, 0.34 per 0.5 unit increase in ETT size; 95 % CI, 0.15-0.79) and with lower peak airway pressure (OR per one cm H2O increment, 0.93; 95 % CI, 0.87-0.99) on multivariate analysis.

Conclusions: Cuff pressure is frequently not maintained within the target range with low-cuff pressure being very common approximately 3 h after routine measurements. Low cuff pressure was associated with lower ETT size and lower peak airway pressure. There is a need to redesign the process for maintaining cuff pressure within the target range.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4607016PMC
http://dx.doi.org/10.1186/s12871-015-0132-7DOI Listing

Publication Analysis

Top Keywords

cuff pressure
44
pressure
14
cuff
13
maintaining cuff
12
target range
12
ett size
12
pressure endotracheal
8
standard practice
8
pressure target
8
cuff pressures
8

Similar Publications

The present study aims to examine the effect of 4 h of continuous sitting on cerebral endothelial function, which is a crucial component of cerebral blood flow regulation. We hypothesized that 4 h of sitting may impair cerebral endothelial function similarly to how it affects lower limb vasculature. Thirteen young, healthy participants were instructed to remain seated for 4 h without moving their lower limbs.

View Article and Find Full Text PDF

Spatiotemporal analysis of the effects of exercise on the hemodynamics of the aorta in hypertensive rats using fluid-structure interaction simulation.

J Transl Int Med

February 2024

Department of Cardiology and Institute of Vascular Medicine, Peking University Third Hospital; State Key Laboratory of Vascular Homeostasis and Remodeling, Peking University; NHC Key Laboratory of Cardiovascular Molecular Biology and Regulatory Peptides, Peking University; Beijing Key Laboratory of Cardiovascular Receptors Research, Beijing 100191, China.

Background And Objective: Hemodynamic changes that lead to increased blood pressure represent the main drivers of organ damage in hypertension. Prolonged increases to blood pressure can lead to vascular remodeling, which also affects vascular hemodynamics during the pathogenesis of hypertension. Exercise is beneficial for relieving hypertension, however the mechanistic link between exercise training and how it influences hemodynamics in the context of hypertension is not well understood.

View Article and Find Full Text PDF

Background: Tranexamic acid (TXA) use has become the gold standard in total joint arthroplasty to limit intraoperative blood loss and transfusion rates. More recently, the indications for TXA have expanded to knee and shoulder arthroscopy with promising early results. However, the effectiveness of TXA during arthroscopic rotator cuff repair (RCR) is unclear.

View Article and Find Full Text PDF

Pocket hematoma is a common and serious complication following cardiac implantable electronic device (CIED) implantation, contributing to significant morbidity and mortality. This study aimed to evaluate the efficacy of a novel pocket compression device in reducing pocket hematoma occurrence. We enrolled 242 patients undergoing CIED implantation, randomly assigning them to receive either the novel compression vest with a pressure cuff or conventional sandbag compression.

View Article and Find Full Text PDF

Noise pollution is a known health risk factor and evidence for cardiovascular diseases associated with traffic noise is growing. At least 20% of the European Union's population lives in noise-polluted areas with exposure levels exceeding the recommended limits of the World Health Organization, which is considered unhealthy by the European Environment Agency. This results in the annual loss of 1.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!