Background: In the context of ultrasound-guided peripheral nerve blocks (regional anesthesia), clear visualization of the needle tip and the target structure are mandatory for the performance of a safe puncture and injection. The purpose of this in vitro study was to analyze the visualization of different forms of needle tips and calibers of cannulas in a phantom simulating human tissue, with the help of a standardized image analysis system. Different depths and angles of the puncture needle in relation to the ultrasound probe were tested.
Material And Methods: Cannula needles established for use in regional anesthesia with different surfaces, diameters and needle tip form in 23 different combinations were analyzed. A gelatine-based phantom was used to simulate human tissue. The standardized visualization of the needle tip with the ultrasound probe was performed in plane at different angles (30°, 45° and 60°), depths of penetration (1 cm, 2 cm and 3 cm) and two different alignments of the cannula needle lumen to the ultrasound probe (0° and 180°). The screenshots of the ultrasound pictures were analyzed by quantifying the pixel brightness around the needle tip (region of interest) with a standardized software (score 0-255).
Results: In this study 1104 ultrasound images of cannula needles were analyzed. Diminished scores (reduced pixel brightness) of the needle tips were documented with increasing distance from the ultrasound probe. Comparison of punctures at angles of 30° and 45° showed no differences in needle tip visibility (same scores) but punctures at an angle of 60° were poorly visualized compared with 30° and 45° (mean scores 87.90 ± 11.60 vs. 78.40 ± 12.07, p < 0.001 and 81.85 ± 11.79 vs. 78.40 ± 12.07, p < 0.001, respectively). The direct alignment of the cannula lumen towards the ultrasound probe (0°) was significantly more easily visualized when compared with the reverse alignment of 180° (mean scores 86.90 ± 12.74 vs. 84.80 ± 11.66, p = 0.003, respectively). No differences in visibility were detected between the different cannula needle diameters examined. The Sprotte cannula showed the best visibility score with respect to the cut of the needle tip (mean score 89.40 ± 11.72).
Conclusion: The visibility of cannulas in ultrasound scans depends on the ultrasound frequency, angle of the puncture in relation to the ultrasound probe and the depth of penetration. The results of this study showed that direct alignment of the cannula needle lumen towards the ultrasound probe (0°) independently improved needle tip visualization. This simple measure allows a significant improvement in the safe performance of ultrasound-guided peripheral nerve blocks.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1007/s00101-015-0098-3 | DOI Listing |
Indian J Crit Care Med
November 2024
Department of Anesthesiology Critical Care, All India Institute of Medical Sciences, Rishikesh, Uttarakhand, India.
Background: To estimate a correlation between change in lung ultrasound aeration score (LUSS) and mortality in patients with ventilator-associated pneumonia.
Materials And Methods: We conducted a prospective observational study in which lung ultrasound, the partial pressure of arterial oxygen to fraction of inspired oxygen ratio (PaO/FiO ratio), and static lung compliance were performed for five consecutive days since the diagnosis of ventilator-associated pneumonia (day 1-5) in a 20-bed multidisciplinary intensive care unit in at a tertiary care academic institute in Northern India. A hundred and seventeen ventilated patients were studied for the first 5 days after ventilator-associated pneumonia (VAP) development.
Int Med Case Rep J
January 2025
Department of Dermatology and Venereology, Faculty of Medicine, Udayana University/Prof. Dr. I.G.N.G Ngoerah General Hospital, Denpasar, Bali, Indonesia.
Gynecomastia is an enlargement of the breasts in men, which is usually benign but can also manifest as a result of malignancies, and can be either unilateral or bilateral. Pseudogynecomastia is male breast enlargement due to excessive fat, while true gynecomastia is a proliferation of glandular tissue. Gynecomastia is common in infants, adolescents, and elderly men, with the most common cause is related to hormonal changes associated with aging.
View Article and Find Full Text PDFNat Commun
January 2025
Department of Biomedical Engineering, Duke University, Durham, NC, USA.
Acoustically probing biological tissues with light or sound, photoacoustic and ultrasound imaging can provide anatomical, functional, and/or molecular information at depths far beyond the optical diffusion limit. However, most photoacoustic and ultrasound imaging systems rely on linear-array transducers with elevational focusing and are limited to two-dimensional imaging with anisotropic resolutions. Here, we present three-dimensional diffractive acoustic tomography (3D-DAT), which uses an off-the-shelf linear-array transducer with single-slit acoustic diffraction.
View Article and Find Full Text PDFJ Med Chem
January 2025
Molecular Imaging Branch, National Institute of Mental Health, National Institutes of Health, Bethesda, Maryland, 20892 United States.
J Clin Gastroenterol
January 2025
The Third Central Hospital of Tianjin, Hedong District.
Goals: To explore dynamic contrast-enhanced ultrasound (CEUS) parameters in predicting hepatic vein pressure gradient (HVPG) for patients with liver cirrhosis (LC).
Background: Noninvasive diagnosis of HVPG remains a challenge.
Study: This prospective study included patients with LC undergoing hepatic vein catheterization and pressure measurement at the hospital from May 2021 to January 2023.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!