Background: Monitoring of an animal's blood glucose concentration is critical for diagnostic and therapeutic decisions. Over the past few decades, portable blood glucose meters (PBGMs) have been used to monitor blood glucose concentrations in animals. Recently, new and improved PBGMs have been made available on the market.
Objective: The purpose of this study was to evaluate four PBGMs for use in dogs and cats.
Animals And Methods: A total of 155 venous blood samples of dogs and 85 venous blood samples of cats were tested using four PBGMs. Control solutions from manufacturers were used to determine the precision of each meter. The coefficient of variation was calculated to determine precision during a set of replicates. Pearson's correlation analysis, Passing-Bablok regression, and Bland-Altman analysis were used to determine the accuracy of four PBGMs against the hexokinase reference method. Error grid analysis was used to evaluate clinical relevance.
Results: All PBGMs, except CERA-PET®, were clinically acceptable for monitoring blood glucose concentrations; AlphaTrak® and VetMate® appeared to be the most accurate ones, demonstrating that to use PBGMs for glucose monitoring, it is important to understand the strengths or limitations of each meter. The difference in results between the PBGMs and the reference method increased at high glucose concentration ranges, which were also affected by the hematocrit.
Conclusions: Although readings of the PBGMs and the reference method varied across glycemic ranges (low, normal, and high glucose concentrations), most PBGMs were clinically acceptable for monitoring blood glucose concentrations in dogs and cats.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1080/01652176.2015.1092617 | DOI Listing |
J Pediatr Nurs
January 2025
Faculty of Nursing, Yarmouk University, Irbid, Jordan. Electronic address:
Background: Type 1 diabetes is the most common endocrine health condition among youth. Healthcare professionals must consider evidence-based guidelines in managing children and adolescents with diabetic ketoacidosis (DKA). The current study aims to assess the outcomes of implementing clinical guidelines by the American Diabetes Association to manage DKA among pediatrics in an emergency department in Palestine.
View Article and Find Full Text PDFMedicine (Baltimore)
January 2025
Department of Clinical Laboratory, Zhejiang Hospital, Hangzhou, Zhejiang, China.
To evaluate the accuracy of home self-monitoring portable blood glucose meters, we analyzed the current problems of patients using portable blood glucose meters and put forward reasonable suggestions. A self-designed questionnaire was used to survey 142 patients and 132 healthcare professionals. The questionnaire consisted of 16 items with an overall score ranging from 1 to 13 (with a higher score indicating better experience).
View Article and Find Full Text PDFJ Med Internet Res
January 2025
Department for Prevention and Care of Diabetes, Faculty of Medicine Carl Gustav Carus, Technische Universität Dresden, Dresden, Germany.
Background: Digital technologies for type 2 diabetes mellitus (T2DM) care hold great potential to improve patients' health in the long term. Only a subset of telemedicine offerings are digital interventions that meet the criteria for prescribable digitale Gesundheitsanwendung (digital health apps; DiGAs) in Germany. Digital treatments further provide vast amounts of patient data that are important to generate evidence.
View Article and Find Full Text PDFPLoS One
January 2025
Faculty of Medicine, University of Kelaniya, Ragama, Sri Lanka.
Background: Cinnamon has been studied as a possible way to control blood glucose and serum cholesterol levels. However, there are no well-conducted randomized controlled trials that can accurately measure the lipid and glucose-lowering effects of Cinnamomum zeylanicum (C. zeylanicum) extract.
View Article and Find Full Text PDFPLoS One
January 2025
Department of Ophthalmology, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, Republic of Korea.
Purpose: We sought to evaluate the relationship between blood vitamin A levels and myopia in adults aged ≥20 years in Korea.
Methods: We collected data of 15,899 participants aged ≥20 years from the Korean National Health and Nutrition Examination Survey. Participants underwent refraction tests to identify myopia and high myopia, and their blood pressure and obesity levels were measured.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!