Paddle-based rotating-shield brachytherapy.

Med Phys

Department of Electrical and Computer Engineering, University of Iowa, 4016 Seamans Center, Iowa City, Iowa 52242 and Department of Radiation Oncology, University of Iowa, 200 Hawkins Drive, Iowa City, Iowa 52242.

Published: October 2015

Purpose: The authors present a novel paddle-based rotating-shield brachytherapy (P-RSBT) method, whose radiation-attenuating shields are formed with a multileaf collimator (MLC), consisting of retractable paddles, to achieve intensity modulation in high-dose-rate brachytherapy.

Methods: Five cervical cancer patients using an intrauterine tandem applicator were considered to assess the potential benefit of the P-RSBT method. The P-RSBT source used was a 50 kV electronic brachytherapy source (Xoft Axxent™). The paddles can be retracted independently to form multiple emission windows around the source for radiation delivery. The MLC was assumed to be rotatable. P-RSBT treatment plans were generated using the asymmetric dose-volume optimization with smoothness control method [Liu et al., Med. Phys. 41(11), 111709 (11pp.) (2014)] with a delivery time constraint, different paddle sizes, and different rotation strides. The number of treatment fractions (fx) was assumed to be five. As brachytherapy is delivered as a boost for cervical cancer, the dose distribution for each case includes the dose from external beam radiotherapy as well, which is 45 Gy in 25 fx. The high-risk clinical target volume (HR-CTV) doses were escalated until the minimum dose to the hottest 2 cm(3) (D(2cm(3)) of either the rectum, sigmoid colon, or bladder reached their tolerance doses of 75, 75, and 90 Gy3, respectively, expressed as equivalent doses in 2 Gy fractions (EQD2 with α/β = 3 Gy).

Results: P-RSBT outperformed the two other RSBT delivery techniques, single-shield RSBT (S-RSBT) and dynamic-shield RSBT (D-RSBT), with a properly selected paddle size. If the paddle size was angled at 60°, the average D90 increases for the delivery plans by P-RSBT on the five cases, compared to S-RSBT, were 2.2, 8.3, 12.6, 11.9, and 9.1 Gy10, respectively, with delivery times of 10, 15, 20, 25, and 30 min/fx. The increases in HR-CTV D90, compared to D-RSBT, were 16.6, 12.9, 7.2, 3.7, and 1.7 Gy10, respectively. P-RSBT HR-CTV D90-values were insensitive to the paddle size for paddles angled at less than 60°. Increasing the paddle angle from 5° to 60° resulted in only a 0.6 Gy10 decrease in HR-CTV D90 on average for five cases when the delivery times were set to 15 min/fx. The HR-CTV D90 decreased to 2.5 and 11.9 Gy10 with paddle angles of 90° and 120°, respectively.

Conclusions: P-RSBT produces treatment plans that are dosimetrically and temporally superior to those of S-RSBT and D-RSBT, although P-RSBT systems may be more mechanically challenging to develop than S-RSBT or D-RSBT. A P-RSBT implementation with 4-6 shield paddles would be sufficient to outperform S-RSBT and D-RSBT if delivery times are constrained to less than 15 min/fx.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC5148177PMC
http://dx.doi.org/10.1118/1.4930807DOI Listing

Publication Analysis

Top Keywords

paddle size
12
delivery times
12
hr-ctv d90
12
s-rsbt d-rsbt
12
p-rsbt
10
paddle-based rotating-shield
8
rotating-shield brachytherapy
8
p-rsbt method
8
cervical cancer
8
treatment plans
8

Similar Publications

Little is known about the diversity and systematics of nematodes from intertidal sands of the Sea of Japan (East Sea). Here, nematodes of the rare genus Gairleanema and two new species of the family Xyalidae are described and illustrated. Metadesmolaimus innii sp.

View Article and Find Full Text PDF

The aim of this study was to analyze the effects of different particle sizes of Diels et Gilg (TDG) powders on physical properties, dissolution, in vitro antioxidant activity, and in vivo hepatoprotective properties. The particle size of TDG coarse powders (TDG-CP), TDG fine powders (TDG-FP), and TDG micro powders (TDG-MP) were measured by a laser particle size analyzer. The physical properties were measured according to the latest version of the Chinese Pharmacopoeia (Committee Chinese Pharmacopoeia 2020).

View Article and Find Full Text PDF

Objective: This study evaluates the clinical applications of the supraclavicular osteocutaneous (SOC) flap in complex head and neck reconstruction and its impact on shoulder function following clavicle resection.

Study Design: Case series with a planned chart review.

Setting: Single-institution tertiary care center.

View Article and Find Full Text PDF

Mechanisms of drug release from a melt-milled, poorly soluble drug substance.

J Pharm Sci

January 2025

TU Dortmund, Department of Biochemical and Chemical Engineering, Laboratory of Solids Process Engineering, Emil-Figge-Str. 68, 44227 Dortmund, Germany. Electronic address:

Increasing the dissolution kinetics of low aqueous soluble drugs is one of the main priorities in drug formulation. New strategies must be developed, which should consider the two main dissolution mechanisms: surface reaction and diffusion. One promising tool is the so-called solid crystal suspension, a solid dispersion consisting of purely crystalline substances.

View Article and Find Full Text PDF

Objectives: To highlight the feasibility of employing super-thin ALT free flaps for reconstruction of complex oral cavity and oropharyngeal defects.

Methods: Retrospective chart review.

Results: Sixteen super-thin ALT free flaps were performed in 15 patients between 2020 and 2023 for reconstruction of oral cavity defects (seven oral tongue, three floor of mouth, three buccal, two oropharynx, one lower lip).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!