CPAP Devices for Emergency Prehospital Use: A Bench Study.

Respir Care

Dipartimento di Scienze Chirurgiche e Diagnostiche Integrate, Sezione Anestesia e Rianimazione, Università degli Studi di Genova, Genova, Italy. Istituto di Ricovero e Cura a Carattere Scientifico (IRCCS) Azienda Ospedaliera Universitaria San Martino-Istituto Scientifico Tumori, Genova, Italy.

Published: December 2015

Background: CPAP is frequently used in prehospital and emergency settings. An air-flow output minimum of 60 L/min and a constant positive pressure are 2 important features for a successful CPAP device. Unlike hospital CPAP devices, which require electricity, CPAP devices for ambulance use need only an oxygen source to function. The aim of the study was to evaluate and compare on a bench model the performance of 3 orofacial mask devices (Ventumask, EasyVent, and Boussignac CPAP system) and 2 helmets (Ventukit and EVE Coulisse) used to apply CPAP in the prehospital setting.

Methods: A static test evaluated air-flow output, positive pressure applied, and FIO2 delivered by each device. A dynamic test assessed airway pressure stability during simulated ventilation. Efficiency of devices was compared based on oxygen flow needed to generate a minimum air flow of 60 L/min at each CPAP setting.

Results: The EasyVent and EVE Coulisse devices delivered significantly higher mean air-flow outputs compared with the Ventumask and Ventukit under all CPAP conditions tested. The Boussignac CPAP system never reached an air-flow output of 60 L/min. The EasyVent had significantly lower pressure excursion than the Ventumask at all CPAP levels, and the EVE Coulisse had lower pressure excursion than the Ventukit at 5, 15, and 20 cm H2O, whereas at 10 cm H2O, no significant difference was observed between the 2 devices. Estimated oxygen consumption was lower for the EasyVent and EVE Coulisse compared with the Ventumask and Ventukit.

Conclusions: Air-flow output, pressure applied, FIO2 delivered, device oxygen consumption, and ability to maintain air flow at 60 L/min differed significantly among the CPAP devices tested. Only the EasyVent and EVE Coulisse achieved the required minimum level of air-flow output needed to ensure an effective therapy under all CPAP conditions.

Download full-text PDF

Source
http://dx.doi.org/10.4187/respcare.04134DOI Listing

Publication Analysis

Top Keywords

air-flow output
20
eve coulisse
20
cpap devices
16
cpap
13
easyvent eve
12
positive pressure
8
boussignac cpap
8
cpap system
8
pressure applied
8
applied fio2
8

Similar Publications

Fluidic Oscillation-Based Pneumatic Actuation for Soft Locomotion and Grasping.

Soft Robot

December 2024

Key Laboratory of Mechanism Theory and Equipment Design of Ministry of Education, School of Mechanical Engineering, Tianjin University, Tianjin, China.

Most pneumatic actuators used in robotics are controlled by valves that contain moving parts (e.g., spool or rotor) and electronics to change the direction or pressure of the air flow.

View Article and Find Full Text PDF

The measurement of airflow velocity is crucial in various fields, and several sensing approaches have been developed for detecting airflow, including optical fiber-based flowmeters. However, these sensors often require complex fabrication processes and precise optical alignment. In this paper, a simpler and more cost-effective approach has been used to measure air flow rate by utilizing the birefringence property of liquid crystals (LCs).

View Article and Find Full Text PDF

Introduction: Biomimetic soft pneumatic actuators (SPA) with Kresling origami patterns have unique advantages over conventional rigid robots, owing to their adaptability and safety.

Objectives: Inspired by cloning and moving behaviors observed from salps, we proposed an SPA based on a Kresling-like pattern with a rigid skeleton. The elongation and output force were tested, and the effectiveness of the applications with the SPA was evaluated.

View Article and Find Full Text PDF

3D Printed Fused Deposition Modeling (FDM) Capillaries for Chemiresistive Gas Sensors.

Sensors (Basel)

July 2023

Department of Food Analysis and Chemistry, Faculty of Technology, Tomas Bata University in Zlin, Vavreckova 5669, 760 01 Zlin, Czech Republic.

This paper discusses the possible use of 3D fused deposition modeling (FDM) to fabricate capillaries for low-cost chemiresistive gas sensors that are often used in various applications. The disadvantage of these sensors is low selectivity, but 3D printed FDM capillaries have the potential to increase their selectivity. Capillaries with 1, 2 and 3 tiers with a length of 1.

View Article and Find Full Text PDF

Research on pressure stability in an air cleaning and supplying system for motorcycle users.

Heliyon

July 2023

Department of Automotive Mechanical Engineering, Faculty of Mechanical Engineering, University of Transport and Communications, Hanoi, Viet Nam.

The article introduces the air purification system arranged on motorcycles to help provide clean air for motorcycle users. This is a new system that is being researched and tested to help reduce the harmful effects of air pollution. First, an air purification system model is proposed and a mathematical model is built.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!