Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Background: The design of femoral component used in total hip arthroplasty is known to influence the incidence of periprosthetic femoral fractures (PFFs) in cementless hip arthroplasty.
Objective: This study was undertaken to determine if 2 potential changes to an existing ABG II-standard cementless implant - addition of a roughened titanium plasma-sprayed proximal coating (ABG II-plasma) and lack of medial scales (ABG II-NMS) could decrease the risk of PFF in the intraoperative and early postoperative periods.
Methods: Six pairs of human cadaveric femurs were harvested and divided into 2 groups, each receiving either of the altered implants and ABG II-standard (control). Each implant was tested in a biomechanical setup in a single-legged stance orientation. Surface strains were measured in intact femurs, during implant insertion, cyclic loading of the bone with the implant, and loading to failure. Strains with the ABG II-standard and the altered implants were compared.
Findings: ABG II-plasma showed better load-bearing capacity, with an average 42% greater failure load than that of ABG II-standard. The cortical hoop, axial and mean strains ABG II-plasma were less than those of ABG II-standard, demonstrating decreased tensile behaviour and better load transfer to the proximal femur. The final residual hoop strains in ABG II-plasma were closer to those of intact bone as compared to the standard stem. No differences in strains were observed between the standard stem and ABG II-NMS.
Conclusion: The increased load-bearing capacity and decreased proximal surface strains on femurs implanted with ABG II-plasma stem should reduce the risks of intraoperative and early postoperative PFF.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.3233/BME-151279 | DOI Listing |