Setting: All 19 public health laboratories in Swaziland that had Xpert(®) MTB/RIF machines installed as part of a countrywide roll-out between June 2011 and June 2014.

Objective: To evaluate the utilisation and functionality of Xpert from 2011 to mid-2014.

Design: Descriptive study of Xpert implementation using routinely collected data.

Results: Of 48 829 Xpert tests conducted, 93% were successful: 14% detected Mycobacterium tuberculosis and 12% showed rifampicin resistance. The most common cause of unsuccessful tests was an 'Error' result (62%). Similar findings were obtained in government-supported and partner-supported laboratories. Annual utilisation of Xpert improved from 51% of maximum capacity in 2011 and 2012 to 74% in 2013 and 2014. A monitoring and supervision exercise of all Xpert testing sites in 2014 showed a generally good performance, with over 50% of laboratories achieving a ⩾80% score on most components. However, poor scores were obtained with equipment use and maintenance (6% achieving a score of ⩾80%), internal audit (19% achieving a score of ⩾80%) and process control (25% achieving a score of ⩾80%).

Conclusion: Countrywide roll-out of Xpert in Swaziland has been successful, although operational issues have been identified and need to be resolved.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4487476PMC
http://dx.doi.org/10.5588/pha.15.0001DOI Listing

Publication Analysis

Top Keywords

countrywide roll-out
12
achieving score
12
xpert® mtb/rif
8
score ⩾80%
8
xpert
6
roll-out xpert®
4
mtb/rif swaziland
4
swaziland three
4
three years
4
years implementation
4