Background: The Xience V USA Study demonstrated safety and efficacy of the XIENCE V(®) everolimus-eluting stent (EES) in a large, prospective study of a real-world, unselected patient population. There is limited long-term data regarding EES performance in high risk patients with bifurcation lesions (BIF). The objective of this analysis was to evaluate the long-term safety and effectiveness of EES in patients with BIF from the XIENCE V USA study.

Methods: The Xience V USA Study was a single arm, prospective, multicenter, real-world study (n = 5,054) undergoing PCI with EES. Baseline data and clinical outcomes at 4 years were evaluated in the subgroup of patients with ≥ 1 BIF who did not undergo a staged procedure. Co-primary endpoints were ARC definite/probable stent thrombosis and a composite of cardiac death and ARC-defined myocardial infarction (MI). Endpoints were adjudicated by an independent CEC.

Results: Of 4,768 patients who did not undergo a staged procedure, there were 511 (10.7%) patients with BIF and 4,257 (89.3%) patients without BIF. Follow-up data was available in 4,459 patients (466 BIF, 3,993 non-BIF). Through binary outcome analysis, at 1 year the overall definite/probable stent thrombosis rates were higher in the BIF group (1.84% vs. 0.76%, P = 0.03). However, at 4 years, the difference in cumulative rates of ARC definite/probable stent thrombosis (BIF 2.3% vs. non-BIF 1.4%, P = 0.13) remained the same as that at 1 year, with no incremental definite/probable stent thrombosis in BIF patients from 2-4 years. The 4-year rates of composite cardiac death and MI were 13.5% for BIF vs. 14.1% for non-BIF (P = 0.78). At 4 years, target lesion failure (19.1% vs. 18.3%, P = 0.66) and ischemia driven-target lesion revascularization (10.2% vs. 10.1%, P = 0.89) were comparable between the two groups.

Conclusions: This subgroup analysis of BIF lesions in a real world population receiving EES demonstrates continued low rates of clinical outcomes in the BIF subgroup at 4 years with no incremental stent thrombosis increase in BIF patients from 2 to 4 years. © 2015 Wiley Periodicals, Inc.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.26217DOI Listing

Publication Analysis

Top Keywords

stent thrombosis
20
xience usa
16
definite/probable stent
16
bif
13
clinical outcomes
12
usa study
12
patients bif
12
patients
10
patients bifurcation
8
bifurcation lesions
8

Similar Publications

Introduction: Reports of pseudoaneurysms associated with biliary self-expandable metallic stent (SEMS) placement have been increasing. Recently, cases of hepatic pseudoaneurysm rupture caused by double pigtail plastic stents (DPS) have also been reported. The symptoms of pseudoaneurysms are often non-specific, and many cases are diagnosed only after rupture.

View Article and Find Full Text PDF

Background: First-generation bioresorbable scaffolds (BRS) increased risks of stent thrombosis and adverse events. The Bioheart scaffold is a new poly-L-lactic acid-based BRS.

Objectives: This study sought to evaluate the efficacy and safety of the BRS in patients with coronary artery disease.

View Article and Find Full Text PDF

Background: The risk-benefit ratio of the Absorb bioresorbable vascular scaffold (BVS) may vary before and after 3 years, the time point of complete bioresorption of the poly-L-lactic acid scaffold.

Objectives: The aim of this study was to determine the time-varying outcomes of the Absorb BVS compared with cobalt-chromium everolimus-eluting stents (EES) from a large individual-patient-data pooled analysis of randomized trials.

Methods: The individual patient data from 5 trials that randomized 5,988 patients undergoing percutaneous coronary intervention to the Absorb BVS vs EES with 5-year follow-up were pooled.

View Article and Find Full Text PDF

Iatrogenic arterial injury is an infrequent but limb-threatening complication of total knee arthroplasty (TKA). Open surgical reconstruction may not always be feasible or optimal, particularly in patients who have recently just undergone complex TKA procedures. In this report, we describe the treatment of a patient who developed popliteal artery occlusion following a complex TKA procedure performed the previous day.

View Article and Find Full Text PDF

Trend, clinical characteristics, and pregnancy outcomes of pregnancy associated venous thromboembolism: a retrospective analysis of nearly 10 years.

J Matern Fetal Neonatal Med

December 2025

Department of Vascular Surgery & Interventional Therapy, Fujian Maternity and Child Health Hospital, College of Clinical Medicine for Obstetrics & Gynecology and Pediatrics, Fujian Medical University, Fuzhou, China.

Background: Pregnancy-associated venous thromboembolism (PA-VTE) seriously threatens maternal health. We aimed to investigate the clinical characteristics, risk factors, treatments, and pregnancy outcomes to better prevent and treat PA-VTE.

Methods: PA-VTE patients were selected from 171,898 women who were registered in the Department of Obstetrics of Fujian Maternity and Child Health Hospital from January 2014 to August 2023 and delivered to calculate the incidence.

View Article and Find Full Text PDF