Goal: Development of a new medical device class generally termed robotic capsule endoscopes (RCE) is currently being pursued by multiple research groups. These maneuverable devices will allow minimally invasive diagnosis and treatment of intestinal pathologies. While the intraluminal pressures related to the migrating motor complex (MMC) are well understood, no previous study has measured the active contact forces exerted by the human small bowel wall on a solid, or near solid bolus such as an RCE. Understanding and quantifying the active contact force are critical for the advancement of RCE technology.

Methods: In this study, the authors develop a novel manometric contact force sensor for human studies and validate the feasibility of the design, sterilization method, and minimally invasive surgical procedure in a multianimal study, followed by a multihuman study.

Results: Four porcine tests of the sensor were conducted. The mean porcine myenteric contact force measured using the new sensor is 1.20 ± 0.08 N·cm. The mean myenteric contact force recorded for all five human test subjects is 0.18 ± 0.33 N·cm.

Conclusion: This study demonstrates the feasibility of operating an MMC force sensor in a live human with a minimally invasive surgical technique and presents force data necessary for RCE design.

Significance: This study represents the first known myenteric contact force measurements on a solid bolus in the human small intestine.

Download full-text PDF

Source
http://dx.doi.org/10.1109/TBME.2015.2479607DOI Listing

Publication Analysis

Top Keywords

contact force
20
force sensor
12
minimally invasive
12
myenteric contact
12
force
8
robotic capsule
8
active contact
8
human small
8
solid bolus
8
invasive surgical
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!