Purpose: To measure and compare the three-dimensional (3D) position of nine different abutments manufactured by different manufacturers after repeated torqueing on an internal-hexagon implant.
Materials And Methods: Nine tapered implants were placed into an acrylic resin block. Five specimens each of nine different abutments (n = 45) were placed into one of nine implants. The abutments were handtightened and then torqued to the manufacturer-recommended torque of 30 Ncm. After 10 minutes, 30 Ncm of torque was reapplied. Another 10 minutes elapsed before testing was completed. Images were recorded in 12-second intervals. The spatial relationship of the abutments to the resin block was determined using 3D digital image correlation. Commercial image correlation software was used to analyze the displacements. Mean displacements for the abutments were calculated in three dimensions and overall for both torque applications. Statistical comparisons were done with a t test and a step-down Bonferroni correction.
Results: The overall 3D displacement of the Atlantis Titanium abutment after the second applied torque was significantly greater than that of two of the eight other abutments. Displacement in all three dimensions for the Atlantis Titanium abutment changed direction between the first and second torque applications. All abutments moved further in the same direction except for the Atlantis Titanium abutment, which moved back toward its original hand-tightened position horizontally after the second torque application.
Conclusion: Re-torqueing of abutments after a 10-minute interval leads to minor displacement of varying degrees between the abutment and a tapered implant. A potential effect of embedment relaxation and/or manufacturing errors should be taken into consideration when selecting an abutment for a cement-retained crown on a tapered implant. Accordingly, clinicians may benefit from adjusting cement-retained implant crowns after re-torqueing the abutments to prevent potential occlusal and interproximal contact problems.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.11607/jomi.3694 | DOI Listing |
J Bone Joint Surg Am
March 2024
School of Medicine, Ponce Health Sciences University, Ponce, Puerto Rico.
Background: Spondylothoracic dysostosis (STD), also known as Jarcho-Levin syndrome (JLS), is a rare autosomal recessive disorder affecting the formation of the spine, characterized by a complete bilateral fusion of the ribs at the costovertebral junction, producing a "crablike" appearance of the thorax. Despite being declared a core indication for a V-osteotomy vertical expandable prosthetic titanium rib (VEPTR) expansion thoracoplasty of the posterior thorax, the natural history of STD in untreated subjects remains poorly documented. In this study, we report radiographic and pulmonary function findings and Patient-Reported Outcomes Measurement Information System (PROMIS) and 24-Item Early Onset Scoliosis Questionnaire (EOSQ-24) scores for untreated adult subjects with STD to gain insights into the natural history.
View Article and Find Full Text PDFPurpose: To evaluate the survival of and incidence of mechanical complications with single- and multiple-unit cement-retained posterior implant-supported restorations with custom CAD/CAM Atlantis titanium abutments (Dentsply Sirona, Charlotte, NC, USA).
Materials And Methods: This retrospective clinical study analysed 196 Astra Tech OsseoSpeed TX Internal Hexagon implants (Dentsply Sirona) placed in 85 patients between January 2011 and January 2021. Customised Atlantis titanium abutments and cement-retained metal-ceramic crowns were employed.
J Prosthet Dent
November 2023
Professor and Program Director, Advanced Education Program in Implant Dentistry, School of Dentistry, Loma Linda University, Loma Linda, Calif.
Statement Of Problem: Screw access channels that emerge on the facial aspect of anterior screw-retained implant crowns can compromise esthetics. Recently, angled screw channels (ASCs) have been developed that can alter screw access channel angulations to improve esthetics. While ASCs can be used on narrow-diameter implants, information is limited on the loss of abutment screw torque on narrow-diameter implants with ASCs.
View Article and Find Full Text PDFJ Prosthodont
January 2024
Division of Dental Biomaterials, Clinic for Reconstructive Dentistry, Center of Dental Medicine, University of Zurich, Zurich, Switzerland.
Purpose: To measure the wear at the implant interface between the Grade 4 titanium (Ti) of the implant and frameworks fabricated using two additively manufactured alloys (Ti alloy and cobalt-chromium [Co-Cr]) pre- and post-artificial aging.
Material And Methods: Three-unit frameworks supported by two implants were additively manufactured (Atlantis; Dentsply Sirona) using Ti and Co-Cr dental alloys. Two implants (OsseoSpeed EV, Astra Tech; Dentsply Sirona) were torqued on each non-engaging framework.
Purpose: To evaluate the clinical, radiographic, and prosthetic outcomes of a single custom CAD/CAM abutment system on implants from four manufacturers, with a mean clinical service of 4 years in a retrospective, multicenter case series study.
Materials And Methods: Adult subjects (n = 142) previously restored with titanium or gold-shaded titanium nitride custom CAD/CAM abutments (Atlantis, Dentsply Sirona; n = 259) connected to implants from four different manufacturers (Dentsply Sirona Implants [AT], Biomet 3i [BM], Nobel Biocare [NB], Straumann [ST]), irrespective of implant-abutment interface, and replacing one or more teeth in any position were recalled for a single-visit examination by calibrated investigators at six university clinics. The primary outcomes evaluated included abutment success and survival; secondary outcomes included assessment of papillae fill, probing pocket depth (PD), bleeding on probing (BOP), marginal bone levels (MBLs), and patient-reported outcome measures (PROMs).