Objectives: To investigate the concept of 'urethral atrophy', which is often cited as a cause of recurrent incontinence after initially successful implantation of an artificial urinary sphincter (AUS); and to investigate the specific cause of the malfunction of the AUS in these patients and address their management.

Patients And Methods: Between January 2006 and May 2013, 50 consecutive patients (mean age 54.3 years) with recurrent incontinence had their AUS explored for malfunction and replaced with a new device with components of exactly the same size, unless there was a particular reason to use something different. Average time to replacement of the device was 10.1 years. The mean follow-up after replacement of the device was 24.7 months. All patients without an obvious cause for their recurrent incontinence had preoperative urodynamic evaluation, including measurement of the Valsalva leak point pressure (VLPP) and the retrograde cuff occlusion pressure (RCOP). After explantation of the AUS in patients without any apparent abnormality of the device at the time of replacement, the pressure generated by the explanted pressure-regulating balloon (PRB) was measured manometrically, when this was possible. In a select group of six consecutive patients of this type, the fibrous capsule surrounding the old cuff was incised then excised to expose and evaluate the underlying corpus spongiosum.

Results: In 31 of the 50 patients (62%) undergoing exploration, a specific cause for the malfunction of their AUS was defined. In the other 19 patients (38%) no cause was found, either preoperatively or at the time of exploration, other than a low VLPP and RCOP. A typical 'waisted' or 'hour-glass' appearance of the underlying corpus spongiosum was demonstrable, to some degree, on explanting the cuff in all cases. In the six patients in whom the restrictive sheath surrounding the cuff was excised, the urethral circumference immediately returned to normal after the compressive effect of the sheath was released. Manometry of the explanted PRBs, when this was possible, showed a loss of pressure in all instances. Replacement of the explanted AUS with a new device with the same size cuff and PRB in 14 of these 19 patients was successful in 12 (85.7%).

Conclusions: These results, and other theoretical considerations, suggest that recurrent incontinence, years after initially successful implantation of an AUS, is because of material failure of the PRB, probably attributable to its age and consequent loss of its ability to generate the pressure it was designed to produce, and that urethral atrophy does not occur. Simply replacing the old device with a new one with the same characteristics, unless there is a particular reason to do otherwise, is usually successful and avoids the complications of alternatives such as as cuff downsizing, implanting a PRB with a higher pressure range, implantation of a second cuff or transcorporeal cuff placement, all of which have been advocated in these patients.

Download full-text PDF

Source
http://dx.doi.org/10.1111/bju.13324DOI Listing

Publication Analysis

Top Keywords

recurrent incontinence
16
patients
10
urethral atrophy
8
implantation artificial
8
artificial urinary
8
urinary sphincter
8
initially successful
8
successful implantation
8
specific malfunction
8
malfunction aus
8

Similar Publications

: Transobturator tape (TOT) procedures are a widely used and effective treatment for stress urinary incontinence (SUI), but there is limited research on mesh-related complications and revision surgeries. This study aimed to evaluate the incidence of revision surgeries and mesh-related complications following TOT procedures and identify potential risk factors influencing these outcomes. : This retrospective study analyzed data from patients who underwent TOT procedures at the specialized incontinence center of University Hospital Aachen (UHA), Germany, between January 2010 and May 2023.

View Article and Find Full Text PDF

Background And Aims: Rectal prolapse (RP) is a debilitating condition and can cause symptoms of fecal incontinence, obstructed defecation, incomplete evacuation of the rectum, and pain. In recent years, there has been increasing evidence that prolapse development is strongly associated with pelvic hiatus (GH) size (normal 4.5 +/- 0.

View Article and Find Full Text PDF

A 54-year-old man presented with gait disturbances, urinary incontinence, and headache for 6 months. Head computed tomography indicated several high-density mass lesions in the quadrigeminal cistern, causing occlusive hydrocephalus. Digital subtraction angiography confirmed tentorial dural arteriovenous fistulae (AVF).

View Article and Find Full Text PDF

The Autologous Fascial Pubovaginal Sling for Recurrent Stress Incontinence: A Retrospective Study.

Low Urin Tract Symptoms

January 2025

Department of Urology, Institute of Urology (Laboratory of Reconstructive Urology), West China Hospital, Sichuan University, Chengdu, Sichuan, People's Republic of China.

Aim: This study aimed to analyze the clinical efficacy and safety of autologous fascial pubovaginal sling (AFPVS) surgery in treating recurrent stress urinary incontinence (SUI) following the failure of mid-urethral sling procedures.

Methods: A retrospective analysis was conducted on the clinical data of SUI patients who underwent AFPVS at our hospital between 2008 June and 2024 June following the failure of mid-urethral sling procedures. The analysis included basic information, surgical parameters, and postoperative complications.

View Article and Find Full Text PDF

Objective: To identify risk factors for recurrence and surgical site infection (SSI) after perineal hernia repair surgery in dogs.

Methods: Medical records of male dogs undergoing perineal herniorrhaphy from 2008 to 2023 at a single institution were retrospectively reviewed. Data pertinent to patient signalment, surgical repair methods, risk factors for SSI, and risk factors for recurrence were collected.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!