Background: The aim of this study was to evaluate the safety and efficacy of the Lumax 740(®) Implantable Cardioverter Defibrillator (ICD) system in patients undergoing a defined 1.5 Tesla (T) MRI.

Materials And Methods: Between November 2013 and April 2014, eighteen patients (age range, 41-78 years; mean age, 64 years) implanted with a Lumax 740(®) ICD system for at least 6 weeks before an MRI were enrolled into this single-center feasibility study. The local ethics committee approved the study before patients gave written informed consent. Patients underwent defined MRI 1.5T of the brain and lower lumbar spine with three safety follow-up evaluations obtained during the 3-month study period. Data were analyzed descriptively. Study endpoints were the absence of either MRI and pacing system related serious adverse device effects (SADE), or of a ventricular pacing threshold increase >0.5V, or of an R-wave amplitude attenuation < 50%, or of an R-wave amplitude < 5.0 mV at 1-month follow-up. The assessment of safety and efficacy was supported by recording of all adverse events, changes in pacing threshold, R-wave sensing, pacing impedances and in battery status.

Results: Sixteen patients completed the MRI and the follow-up period. As no SADE occurred, the SADE free rate was 100%. Freedom from ventricular pacing threshold increase was 100% (16/16; 95%CI: 82.9%; 100.0%). There were no significant differences between baseline and follow-up measurements of sensing amplitudes (-0.58 ± 2.07 mV, P = 0.239, -0.41 ± 1.04 mV, P = 0.133, and -0.25 ± 1.36 mV, P = 0.724, for immediately after, 1 month and 3 months after MRI scan, respectively) and pacing thresholds (-0.047 ± 0.18 V, P = 0.317, -0.019 ± 0.11 V, P = 0.490, and 0.075 ± 0.19 V, P = 0.070, for immediately after, 1 month and 3 months after MRI scan, respectively). Lead impedances after the MRI scan were significantly lower as compared with baseline values (-22.8 ± 21.69 Ω, P = 0.001, -21.62 ± 39.71 Ω, P = 0.040, and -33.68 ± 57.73 Ω, P = 0.018, for immediately after, 1 month and 3 months after MRI scan, respectively).

Conclusion: MRI scans in patients with MRI conditional ICD system (Lumax 740(®) ) are feasible and can be performed safely under defined conditions in a hospital setting.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jmri.25037DOI Listing

Publication Analysis

Top Keywords

implantable cardioverter
8
cardioverter defibrillator
8
feasibility study
8
lumax 740®
8
icd system
8
study
6
clinical safety
4
mri
4
safety mri
4
mri conditional
4

Similar Publications

Background: Nurses provide essential care for symptomatic chronic Chagas disease carriers, caused by Trypanosoma cruzi, offering crucial support, symptom management, medication administration, and monitoring to enhance their health-related quality of life.

Objective: To increase healthcare professionals' awareness of the critical role played by high-quality care in the management of patients with chronic Chagas disease.

Methods: This scoping review employed the PRISMA-ScR method as a framework for article selection.

View Article and Find Full Text PDF

[Remote monitoring for patients with cardiac implantable electronic devices and heart failure].

Herzschrittmacherther Elektrophysiol

January 2025

Hannover Herzrhythmus Centrum, Klinik für Kardiologie und Angiologie, Medizinische Hochschule Hannover, Carl-Neuberg-Str. 1, 30625, Hannover, Deutschland.

The digitalization in healthcare facilitates continuous monitoring of relevant medical parameters through internal and external sensors. For patients with heart failure and cardiac implantable electronic devices (CIEDs), telemedicine has the potential to improve patient care and reduce use of healthcare resources. Remote monitoring shortens the response time to a clinical event, reduces inappropriate shocks, and increases patient satisfaction.

View Article and Find Full Text PDF

Congenital heart disease (CHD) is the most common congenital anomaly in newborns. Advances in catheter and surgical techniques led to the majority of these patients surviving into adulthood, leading to evolving challenges due to the emergence of long-term complications such as arrhythmias. Interventional electrophysiology (EP) has had remarkable advances over the last few decades, and various techniques and devices have been explored to treat adult patients with CHD.

View Article and Find Full Text PDF

Despite arrhythmogenic right ventricular cardiomyopathy (ARVC) being predominantly a right ventricular (RV) disease, concomitant left ventricular (LV) involvement has been recognized. ARVC is diagnosed by the RV-centric 2010 Task Force Criteria(TFC) using routine echocardiography, but previous studies have suggested that strain imaging may be more sensitive to detect RV and LV dysfunction. No data however are available regarding the additional value of combining biventricular strain for risk stratification.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!