A Prototype Detector for a Novel High-Resolution PET System: BazookaPET.

IEEE Nucl Sci Symp Conf Rec (1997)

Department of Medical Imaging, University of Arizona, Tucson, AZ, 85724 USA. They are also with the College of Optical Sciences, University of Arizona, Tucson, AZ 85721 USA.

Published: January 2012

We have designed and are developing a novel proof-of-concept PET system called BazookaPET. In order to complete the PET configuration, at least two detector elements are required to detect positron-electron annihilation events. Each detector element of the BazookaPET has two independent data acquisition channels. One side of the scintillation crystal is optically coupled to a 4×4 silicon photomultiplier (SiPM) array and the other side is a CCD-based gamma camera. Using these two separate channels, we can obtain data with high energy, temporal and spatial resolution data by associating the data outputs via several maximum-likelihood estimation (MLE) steps. In this work, we present the concept of the system and the prototype detector element. We focus on characterizing individual detector channels, and initial experimental calibration results are shown along with preliminary performance-evaluation results. We measured energy resolution and the integrated traces of the slit-beam images from both detector channel outputs. A photo-peak energy resolution of ~5.3% FWHM was obtained from the SiPM and ~48% FWHM from the CCD at 662 keV. We assumed SiPM signals follow Gaussian statistics and estimated the 2D interaction position using MLE. Based on our the calibration experiments, we computed the Cramér-Rao bound (CRB) for the SiPM detector channel and found that the CRB resolution is better than 1 mm in the center of the crystal.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4548890PMC
http://dx.doi.org/10.1109/NSSMIC.2012.6551486DOI Listing

Publication Analysis

Top Keywords

prototype detector
8
pet system
8
detector element
8
energy resolution
8
detector channel
8
detector
6
detector novel
4
novel high-resolution
4
high-resolution pet
4
system bazookapet
4

Similar Publications

CoReSi: a GPU-based software for Compton camera reconstruction and simulation in collimator-free SPECT.

Phys Med Biol

January 2025

CREATIS, INSA de Lyon, Bâtiment Blaise Pascal, 7 Avenue Jean Capelle, Villeurbanne, 69621 Cedex , FRANCE.

Compton cameras are imaging devices that may improve observation of sources of γ photons. We present CoReSi, a Compton Reconstruction and Simulation software implemented in Python and powered by PyTorch to leverage multi-threading and for easy interfacing with image processing and deep learning algorithms. The code is mainly dedicated to medical imaging and for near-field experiments where the images are reconstructed in 3D.

View Article and Find Full Text PDF

Objective: Patient positioning during clinical practice can be challenging, and mispositioning leads to a change in CT number. CT number fluctuation was assessed in single-energy (SE) EID, dual-energy (DE) EID, and deep silicon photon-counting detector (PCD) CT over water-equivalent diameter (WED) with different mispositions.

Methods: A phantom containing five clinically relevant inserts (Mercury Phantom, Gammex) was scanned on a clinical EID CT and a deep silicon PCD CT prototype at vertical positions of 0, 4, 8, and 12 cm.

View Article and Find Full Text PDF

Defect detection and classification in super-high reflector mirrors and their substrates are crucial for manufacturing laser gyroscope systems. This paper presents a prototype designed to meet the requirements for the reflection and transmission of laser gyroscope mirror substrates. The prototype featured two measurement channels (bright field and dark field) and could detect defects on patterned and unpatterned surfaces.

View Article and Find Full Text PDF
Article Synopsis
  • PET imaging using Zr is gaining traction for visualizing immune and cancer cells, but has limitations due to low image quality from gamma ray emissions.
  • A new prototype has been developed to improve Compton imaging within PET by redesigning the scatter detector and optimizing image reconstruction methods.
  • The upgraded WGI system showed enhanced imaging performance, including improved angular resolution, sensitivity, and reduced noise, demonstrating successful in-vivo imaging of a mouse with tumor burden after Zr labeled antibody injection.
View Article and Find Full Text PDF

Synchrotron microbeam radiotherapy (MRT), which has entered the clinical transfer phase, requires the development of appropriate quality assurance (QA) tools due to very high dose rates and spatial hyperfractionation. A microstrip plastic scintillating detector system with associated modules was proposed in the context of real-time MRT QA. A prototype of such a system with 105 scintillating microstrips was developed and tested under MRT conditions.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!