Biomechanics of Corneal Ring Implants.

Cornea

*Gutsehen Eye Center, Ybbs, Austria; and †Department of Ophthalmology, Medical University of Innsbruck, Innsbruck, Austria.

Published: November 2015

Purpose: To evaluate the biomechanics of corneal ring implants by providing a related mathematical theory and biomechanical model for the treatment of myopia and keratoconus.

Methods: The spherical dome model considers the inhomogeneity of the tunica of the eye, dimensions of the cornea, lamellar structure of the corneal stroma, and asphericity of the cornea. It is used in this study for calculating a strengthening factor sf for the characterization of different ring-shaped corneal implant designs. The strengthening factor is a measure of the amount of strengthening of the cornea induced by the implant.

Results: For ring segments and incomplete rings, sf = 1.0, which indicates that these implants are not able to strengthen the cornea. The intracorneal continuous complete ring (MyoRing) has a strengthening factor of up to sf = 3.2. The MyoRing is, therefore, able to strengthen the cornea significantly.

Conclusions: The result of the presented biomechanical analysis of different ring-shaped corneal implant designs can explain the different postoperative clinical results of different implant types in myopia and keratoconus.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC4598073PMC
http://dx.doi.org/10.1097/ICO.0000000000000591DOI Listing

Publication Analysis

Top Keywords

strengthening factor
12
biomechanics corneal
8
corneal ring
8
ring implants
8
ring-shaped corneal
8
corneal implant
8
implant designs
8
strengthen cornea
8
cornea
5
ring
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!